Siderail Operation Guide
18
POSITIONING SIDERAILS
NOTE
The siderails can be locked at two heights (intermediate & full up). The foot end siderails on a Short Bed do
not have an intermediate position.
The siderails can be tucked away under the bed when not in use. To remove the rail from the tucked position,
grasp the top of the rail and pull outward.
To engage the head end siderail, grasp the rail and swing it upward to full height. When the siderail is being
raised, it does not lock in the intermediate position. To lower the siderail, push in the blue release handle (A)
and rotate the siderail until it locks in the intermediate position. To lower the siderail fully, push in the blue
release handle (A) again and rotate the siderail until it is completely lowered.
To engage the foot end siderail, the same procedure is required as for the head end siderail, however, the
siderail swings toward the foot end of the bed.
WARNING
Be sure the siderail is locked se-
curely into position. Siderails are
not intended to keep patients
from exiting the bed. They are
designed to keep a patient
from inadvertently rolling off
the bed. Proper restraint
methods should be utilized to
ensure the patient remains in bed.
The siderails are not intended to be
used as a push device.
To disengage the rail, push in the blue release handle (A) and swing the rail down to the desired height. Store
the siderails by tucking the rails under the bed. Rails must be in the full down position before they can be
tucked.
SIDERAIL CONTROL PANEL LIGHTS
The bed is equipped with lights to illuminate the head end siderail control panel and the red nurse call
switches. Both can be activated at the foot board control panel. Three settings are available for the control
panel lights: low, medium and high intensity. When all lights are off, push the siderail control light button at
the foot board once to turn on both the control lights and the nurse call light at the siderail. Push again to
change from low to medium setting, and a third time to change to the high setting. The nurse call light intensity
is not affected. Pushing the button a fourth time will turn off the siderail control panel lights and pushing it
a fifth time will turn off the red nurse call light as well (see control panel guide
NOTE
The purpose of the red nurse call light on the siderails is to ensure the patient immediately knows which button
to push to contact the nurse station. Turning the red light off may compromise this ability, especially in a dark-
ened room.
A
Содержание Secure II 3002
Страница 33: ...Optional Pendant Operation 32 ...
Страница 46: ...Bed Circuit Boards 45 CPU BOARD 3002 407 950 ...
Страница 70: ...Service Information 69 POWER AND SENSOR COIL CORD REPLACEMENT ILLUSTRATION A B A B VIEW FROM CENTER OF BED ...
Страница 98: ...Notes 97 ...
Страница 101: ...Base Assembly and Options 100 FOOT END HEAD END ...
Страница 102: ...Base Assembly and Options 101 Installation of Roll Pin into Brake Shaft Crank Typical 4 Places ...
Страница 103: ...Base Assembly and Options 102 HEAD END ...
Страница 104: ...Base Assembly and Options 103 HEAD END LEFT SIDE FOOT END ...
Страница 105: ...Base Assembly and Options 104 Zoom Option Zoom Option Steer Option ...
Страница 106: ...Base Assembly and Options 105 Typical Both Ends Typical Both Ends ...
Страница 110: ...109 Lift Assembly Common Assembly part number 3000 200 275 reference only ...
Страница 124: ...Notes 121 2 ...
Страница 125: ...Base Assembly with Zoomr Option 122 HEAD END ...
Страница 126: ...Base Assembly with Zoomr Option 123 HEAD END HEAD END 3001 200 306 See Detail A 3001 200 306 DETAIL A ...
Страница 127: ...Base Assembly with Zoomr Option 124 HEAD END HEAD END ...
Страница 128: ...Base Assembly with Zoomr Option 125 FOOT END 8 Caster Option 6 Caster Option ...
Страница 129: ...Base Assembly with Zoomr Option 126 FOOT END HEAD END ...
Страница 130: ...Base Assembly with Zoomr Option 127 ...
Страница 131: ...Base Assembly with Zoomr Option 128 ...
Страница 138: ...Base Assembly with Battery Back Up Option 135 Assembly part number 3002 350 170 reference only HEAD END HEAD END ...
Страница 139: ...Base Assembly with Battery Back Up Option 136 HEAD END HEAD END 2025 31 805 Ground Strap ...
Страница 140: ...Base Assembly with Battery Back Up Option 137 ...
Страница 141: ...Base Assembly with Battery Back Up Option 138 ...
Страница 144: ...Litter Assembly 141 ...
Страница 145: ...Litter Assembly 142 ...
Страница 146: ...Litter Assembly 143 ...
Страница 147: ...Litter Assembly 144 DETAIL A Standard Length Bed 8 Shorter Length Bed ...
Страница 149: ...Litter Assembly 146 DETAIL F ...
Страница 153: ...Electrical Litter Assembly 150 ...
Страница 154: ...Electrical Litter Assembly 151 Siderail Cable Attachment ...
Страница 155: ...Electrical Litter Assembly 152 DETAIL C To Gatch Motor See Detail C ...
Страница 156: ...Electrical Litter Assembly 153 DETAIL A DETAIL B 115V BED 230V BED ...
Страница 157: ...Electrical Litter Assembly 154 SMART TV OPTION UNITED KINGDOM POWER CORD CABLE INSTALLATION ...
Страница 159: ...Mechanical Litter Assembly 156 ...
Страница 160: ...Mechanical Litter Assembly 157 ...
Страница 161: ...Mechanical Litter Assembly 158 Standard Length Bed Fowler Drive 8 Shorter Bed Option Fowler Drive ...
Страница 162: ...Mechanical Litter Assembly 159 SECTION A A A A ...
Страница 164: ...Notes 161 ...
Страница 169: ...166 Fowler Drive Assembly Standard Fowler Drive Shown ...
Страница 178: ...Notes 175 ...
Страница 180: ...Optional Scale Bed Exit Assembly 177 FOOT END ...
Страница 182: ...Optional HWI Assembly with Communication 179 ...
Страница 183: ...Optional HWI Assembly w Communication Ports 180 ...
Страница 184: ...Optional HWI Assembly w Communication Ports 181 ...
Страница 187: ...Zoom Option Litter Assembly 184 DETAIL A Assembly part number 3002 131 10 reference only ...
Страница 188: ...Zoom Option Litter Assembly 185 DISPLAY HOUSING ...
Страница 189: ...Zoom Option Litter Assembly 186 On Off Switch Inside Enclosure ...
Страница 190: ...Zoom Option Litter Assembly 187 ...
Страница 191: ...Zoom Option Litter Assembly 188 ...
Страница 194: ...Notes 191 ...
Страница 198: ...Battery Back Up Option Head End Enclosure 195 ...
Страница 200: ...Head End Siderail Assembly 197 LEFT SIDERAIL SHOWN ...
Страница 201: ...Head End Siderail Assembly 198 CARRIER ARM ASSEMBLY LEFT SIDERAIL SHOWN ...
Страница 202: ...Head End Siderail Assembly 199 SIDERAIL OUTER PANEL ASSEMBLY 8 SHORTER BED ONLY RIGHT AND LEFT SIDERAILS ...
Страница 203: ...Head End Siderail Assembly 200 SIDERAIL INNER PANEL ASSEMBLY ...
Страница 204: ...Head End Siderail Assembly 201 LEFT HEAD END SIDERAIL WIRE ROUTING ...
Страница 205: ...Head End Siderail Assembly 202 RIGHT HEAD END SIDERAIL WIRE ROUTING ...
Страница 223: ...3002 401 300 Foot End Siderail Assembly Left 3002 401 400 Foot End Siderail Assembly Right 220 LEFT SIDERAIL SHOWN ...
Страница 257: ...2040 120 4 Optional Pleur Evac Rack w Defib Tray 252 ...
Страница 263: ...258 3001 333 000 Optional Accessory Adapter Frame ...
Страница 265: ...3002 335 Optional Litter Roller Assembly 260 Item Part No Part Name Qty A page 261 Roller Assembly 2 ...
Страница 268: ...3001 318 20 Optional Bed Extender Assembly 263 C F H C D C D E J ...
Страница 269: ...3000 318 50 Optional Bed Extender Mattress 264 ...