Do not clean, disinfect, service, or perform maintenance while the product is in use.
•
Always unplug the power cord from the wall outlet when large spills occur near the circuit boards and cables. Clean up
the fluid, and inspect the product. Fluids can cause unpredictable operation and decreased functionality of any electrical
product. Do not return the product to service until dry and tested for safe operation.
•
Always wipe down with clean water (or 70% isopropyl alcohol, if using V
Viirre
exx® TB) and dry each product after
disinfecting. Some disinfectants are corrosive in nature and may cause damage to the product. If you do not rinse and
dry the product, you may leave a corrosive residue on the surface of the product. This corrosive residue could cause
premature degradation of critical components. Failure to follow these disinfecting instructions may void your warranty.
Recommended disinfectants for this product’s surfaces include:
•
Quaternary (active ingredient - ammonium chloride)
•
Phenolic (active ingredient - o-phenylphenol)
•
Chlorinated bleach solution (10,000 ppm available chlorine, 941 mL of a 5.25% sodium hypochlorite solution per 4000
mL of water)
•
Alcohol (active ingredient - 70% isopropyl alcohol)
•
Accelerated hydrogen peroxide (5,000 ppm hydrogen peroxide)
Disinfection method:
1. Follow the disinfectant solution manufacturer’s dilution recommendations.
2. Using spray or pre-soaked wipes, apply the recommended disinfectant solution.
3. Hand wash all exposed surfaces of the product with the recommended disinfectant.
4. Dry the product before you return it to service.
N
No
otte
e
•
Avoid oversaturation. Do not allow the product to remain wet.
•
Follow the manufacturer’s dilution recommendations for appropriate contact time and rinse requirements. Follow the
chemical manufacturer’s guidelines to disinfect.
5212-009-201 Rev A.0
15
EN
Содержание Secure Connect Connected Hospital 5212
Страница 2: ......
Страница 4: ......
Страница 25: ...Torque item AA to 0 29 0 05 ft lb Torque item AB to 0 34 0 05 ft lb 5212 009 201 Rev A 0 21 EN ...
Страница 26: ...Torque item AC to 0 58 0 05 ft lb EN 22 5212 009 201 Rev A 0 ...
Страница 30: ...Torque item AA to 0 29 0 05 ft lb Torque item AB to 0 34 0 05 ft lb EN 26 5212 009 201 Rev A 0 ...
Страница 31: ...Torque item AC to 0 58 0 05 ft lb 5212 009 201 Rev A 0 27 EN ...
Страница 43: ......
Страница 44: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 5212 009 201 Rev A 0 WCR AA 14 2020 10 ...