Operation
i
Bed Awareness (option) (Continued)
Configuring
i
Bed Awareness
WARNING
•
Do not use
i
Bed
Awareness to replace your patient monitoring protocol.
•
Do not use
i
Bed
Awareness as a lock indicator for siderails.
i
Bed
Awareness only detects the position of the
siderails.
CAUTION
•
Always make sure that the siderails are locked before you arm
i
Bed
Awareness.
•
Make sure that you set the desired product parameters before arming
i
Bed
Awareness.
•
Do not use accessories that cover the footboard LED light bars.
When enabled,
i
Bed
Awareness helps to monitor
S3
status and parameter conditions.
To monitor a parameter, place the product to the desired position. You can monitor the low height position,
Chaperone
bed exit, and Fowler 30°+.
When armed,
i
Bed
Awareness automatically monitors all current siderail positions and the brake.
To arm
i
Bed
Awareness, press
i
Bed
(A) (see
).
Note:
If there is an error in one of the product functions, an error code will appear.
i
Bed
Awareness will not arm. For
more information about error codes, see the maintenance manual.
After you arm
i
Bed
Awareness, the LED light bars on the footboard illuminate green and the monitored footboard LED
indicators on the footboard illuminate.
To set an alarm tone for
i
Bed
Awareness, see
To disarm
i
Bed
Awareness, press
i
Bed
again (A) (see
).
Note:
The settings for lockout controls, scale calibration data, bed exit, and
i
Bed
Awareness are preserved when the
product is unplugged, or during a power failure.
Acknowledging
i
Bed Awareness status alerts
If the parameter conditions selected for
i
Bed
Awareness are changed:
•
LED light bars on the footboard flash amber
•
changed indicator LED on the footboard LED indicator flashes
•
sound alarm is triggered
•
changed parameter status alert is displayed on the display screen