Setup
Equipping the optional nurse call communication (Continued)
To activate the optional nurse call communication, see
Activating nurse call communication (option) on page 30
.
A
B
Figure 3: Nurse call pendant port and 37-pin connector
Setting up the optional Zoom
If your product is equipped with the optional
Zoom®
drive wheel option, follow the setup instructions below before
putting the product into service:
1. Plug the product into a grounded, hospital grade wall outlet. The 12V batteries that provide power to the drive wheel
and the backup power to the unit functions, will charge whenever the power cord is plugged into a wall outlet.
Note:
You must charge the batteries for approximately 10 hours before the product is put into service.
2. Unplug the power cord from the wall outlet. Push the battery power switch located on the lower left corner of the
head end to the on position (
Note:
Make sure that each function is working on the footboard control panel and the head end siderails (see
).
3. With the battery power switch on and the brakes applied, make sure that the
Release brakes
LED illuminates on the
head end control panel (
motorized drive panel (option) on page 59
).
4. With the battery power switch on and the drive wheel in the neutral position, make sure that the
Engage drive wheel
LED illuminates on the head end control panel (