Operating a Light Source
Please note the following special warnings to avoid user or patient injury or
product damage when using a light source.
IMPORTANT SAFETY NOTICE - HIGH TEMPERATURES:
Before operating this device, please read this operating manual
thoroughly and carefully. When using a light source, fire and/or
severe injury may result to the patient, user or inanimate objects
if the instructions in this manual are not followed.
All light sources can generate significant amounts of heat
(exceeding 41 °C/106 °F) at the scope tip, the scope light post,
the light cable tip, and/or near the light cable adapter. Higher
levels of brightness result in higher levels of heat. Adjust the
brightness level of the camera and the monitor to the minimum
brightness necessary to adequately illuminate the surgical site.
In addition, adjust the internal shutter of the camera higher in
order to run the light source at a lower intensity. Avoid touching
the scope tip or the light cable tip to the patient, and never
place them on top of the patient, as doing so may result in burns
to the patient or user. In addition, never place the scope tip, the
scope light post, the light cable adapter, or the light cable tip
on the surgical drapes or other flammable material, as doing so
may result in fire.
Always place the light source in Standby mode and disable ENV
mode before the scope is removed from the light cable or the
device is unattended. The scope tip, scope light post, light cable
adapter, and light cable tip will take several minutes to cool off
after being placed in Standby mode, and therefore may still
result in fire or burns to the patient, user, or inanimate objects.
The warranty is void if any of the above warnings or cautions are disregarded.