13. Verify proper operation before you return the product to service.
H
Ho
oo
okk rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
Stubby #2 Phillips screwdriver
•
Pick
P
Prro
occe
ed
du
urre
e::
1. Unplug the pump from AC power.
2. Unplug the mattress tubing from the pump.
3. Place the pump front down on a work surface.
4. Using a pick, remove and discard the information label on the hook assembly.
5. Using a #2 Phillips, remove the screw (under the label removed in step 4) that secure the hook assembly to the back
pump housing.
6. Using a T20 Torx driver, remove and save the six screws that secure the back pump housing to the front pump housing.
7. Using two hands, rotate the pump assembly to the right so the front is to your left.
8. Separate the front and back pump housing. Place the back pump housing down on the work surface.
9. Unplug the power input cable and the power output cable from the power supply.
10. Using one hand to hold the fish paper and a #2 Phillips screwdriver, remove and save the four Phillips screws and
washers that secure the power supply to the main frame.
11. Remove the power supply to the side and fish paper.
12. Using a #2 Phillips screwdriver, remove and save the four screws that secure the pump frame to the main frame and set
aside.
13. Using a #2 Phillips screwdriver remove and save the six screws that secure the hook assembly to the back pump
housing.
14. Remove and discard the hook assembly.
15. Reverse the steps to reinstall (
16. Run and pass the diagnostic test.
17. Verify proper operation before you return the product to service.
F
Frro
on
ntt p
pu
um
mp
p h
ho
ou
ussiin
ng
g rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
#2 Phillips screwdriver
•
Stubby #2 Phillips screwdriver
P
Prro
occe
ed
du
urre
e::
1. Unplug the pump from AC power.
2. Unplug the mattress tubing from the pump.
3. Place the pump front down on a work surface.
4. Using a T20 Torx driver, remove and save the six screws that secure the back pump housing to the front pump housing.
5. Using two hands, rotate the pump assembly to the right so the front is to your left.
2874-009-002 Rev A.1
11
EN