N
•
Always use ESD protective equipment before opening antistatic bags and servicing electronic parts.
•
Do not place unprotected circuit boards on the floor.
N
No
otte
e -- Always ship back circuit boards to Stryker in the same antistatic bags that the new boards were originally shipped in.
The electronic circuits in the product are completely protected from static electricity damage when factory assembled.
Always use adequate static protection when servicing the electronic systems of the product. All service personnel must use
static protection whenever they are touching wires.
Sample antistatic protection equipment includes:
•
1 antistatic wrist strap
•
1 grounding plug
•
1 test lead with a banana plug on one end and an alligator clip on the other end
Make sure that you follow the ESD manufacturer’s instructions for appropriate protection against static discharge.
P
Po
ow
we
erriin
ng
g o
offff tth
he
e p
pu
um
mp
p
Press the power button and hold for 2 seconds to power off the pump.
N
No
otte
e
•
Allow the support surface to deflate before you unplug the power cord.
•
For optional deflation, remove the CPR plug from the support surface.
F
Fu
usse
e rre
ep
plla
acce
em
me
en
ntt
Fuse rating: F2AH250V
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Flat head screwdriver
P
Prro
occe
ed
du
urre
e::
1. Unplug the pump from AC power.
2. Unplug the mattress tubing from the pump.
3. Place the pump front down on a work surface.
4. Using a flat head screwdriver, remove and save the fuse cover located below the power inlet.
5. Remove and dispose of the fuse.
6. Reverse the steps to reinstall.
7. Verify proper operation before you return the product to service.
B
Bu
um
mp
pe
err rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Pick
EN
8
2874-009-002 Rev A.1
Содержание IsoTour
Страница 2: ......
Страница 6: ...EMC information 35 EN 2 2874 009 002 Rev A 1 ...
Страница 43: ......
Страница 44: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA 2874 009 002 Rev A 1 WCR AA 21 2019 08 ...