Страница 1: ...Medical For parts or technical assistance call 800 327 0770 option 2 Epic II Critical Care Bed Model 2030 ...
Страница 2: ...he Bed 22 Adding or Removing Items During a Patient s Stay 23 Displaying Trendelenburg or Fowler Angle 24 Converting the Patient s Weight 24 Displaying the Weight Log 25 Viewing the Patient s Weight in Gain Loss Mode 26 Changing the Numerical Value of Displayed Weight 27 Optional Pendant Operation 28 Optional Epic II Head End Control Panel Operation 29 Battery Charging and Operation Epic II Option...
Страница 3: ... Bed Length Width L 91 W 42 5 or L 231 cm W 108 cm Minimum Maximum Bed Height Standard Minimum Maximum Bed Height Enhanced 18 to 32 5 0 5 46 cm to 82 5 cm 19 5 to 34 5 0 5 49 5 cm to 88 cm Add 2 inches if the bed has 8 casters Fluoro Access 17 5 Epic II 16 Epic II Knee Gatch Angle 0 to 30 Back Angle 0 to 90 Trendelenburg Reverse Trendelenburg 12 to 12 2 Electrical Requirements 115 VAC 60 Hz 7 0 Am...
Страница 4: ...Bed is equipped with a hospital grade plug for protection against shock hazard It must be plugged directly into a properly grounded three prong receptacle Grounding reliability can be achieved only when a hospital grade receptacle is used Serious injury can result if caution is not used when operating the bed Operate bed only when all persons are clear of the electrical and mechanical systems Leav...
Страница 5: ...n working under the bed always place blocks under the litter frame to prevent injury in case the Bed Down switch is accidently activated The battery tray assembly weighs 50 pounds Take care when removing the two hex head screws secur ing it to the base frame or personal injury could result Battery posts terminals and related accessories contain lead and lead compounds chemicals known to the State ...
Страница 6: ...lure caused by fluids could even cause the bed to operate unpredictably and could cause injury to the patient DO NOT put the bed back into service until it is completely dry and has been thoroughly tested for safe operation Preventative maintenance should be performed at a minimum of annually to ensure all features are func tioning as designed Close attention should be given to safety features inc...
Страница 7: ... the up and intermediate positions Run through each function on the foot board control panel and ensure that each is working properly see function lockout system usage page 15 and scale system control panel guide page 20 Ensure all functions are working properly on the siderail controls Raise the Back up to approximately 60 Squeeze the CPR release handle and ensure the Back and Knee will drop with...
Страница 8: ...caution in that means are provided for the connection of the EQUIPMENT to the protective earth con ductor in the fixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become live in the event of a failure of the BASIC INSULATION Mode of Operation Continuous IPX4 Protection from liquid splash Dangerous Voltage Symbol Protective Earth Terminal Potential Equalization Symbo...
Страница 9: ...to the upper position NOTE There are LED lights on the outside of the head end siderails that will blink when the brakes are not engaged only if the bed is plugged into a wall socket see page 11 The brakes will still operate properly when the bed is not plugged in STEER PEDAL OPERATION The purpose of the steer caster is to help guide the bed along a straight line and to help with pivoting at corne...
Страница 10: ...n 80 when the CPR emergency release is activated Attempting to lower the Back in this position without assistance may result in injury to the operator CPR BOARD USAGE The CPR board is stored on the bed s head board To remove it pull it away from the head board and lift it out of the storage position The head board can also be removed and used as an emergency CPR board FOLEY BAG HOOKS USAGE The sta...
Страница 11: ...red or lowered in Reverse Trendelenburg the siderails tucked under the litter in the storage position and the casters turned there is the po tential for interference between the siderail and the caster Raise the siderails when lowering the litter to the full down position to prevent the interference from causing the bed s scale system to weigh inaccurately The siderails are not intended to be used...
Страница 12: ...8 Press to raise the litter If your bed is equipped with the enhanced height option continue to hold the button an additional 5 seconds after the first stop The litter will raise an additional 2 inches 9 Press to lower the litter 10 Press to activate emergency CPR positioning 11 Press to activate Cardiac Chair positioning CAUTION When attaching equipment to the bed ensure it will not impede normal...
Страница 13: ... Press to turn on the TV or radio Press again to change TV channels and to turn off the TV 7 Press to increase the TV or radio volume 8 Press to decrease the TV or radio volume 9 Press to turn on the room lights Press again to turn off 10 Press to turn on the reading light Press again to turn off CAUTION When attaching equipment to the bed ensure it will not impede normal bed operation For example...
Страница 14: ...e TV or radio volume 8 Press to decrease the TV or radio volume 9 Press to turn on the room lights Press again to turn off 10 Press to turn on the reading light Press again to turn off 11 Press to change the TV channel up 12 Press to change the TV channel down 13 Press to mute the TV sound Press again to turn the sound back on 14 Press to display TV closed captioning CAUTION When attaching equipme...
Страница 15: ...ls Press again to unlock 6 Press to raise the bed height If your bed is equipped with the enhanced height option continue to hold the button an additional 5 seconds after the first stop The litter will raise an additional 2 inches 7 Press to lower bed 8 Press to lower head end of bed Trendelenburg 9 Press to lower foot end of bed Reverse Trendelenburg CAUTION When attaching equipment to the bed en...
Страница 16: ...be lighted when that function is locked out 2 To lock out the bed movement functions on the siderails and prevent the patient from changing the posi tioning of the bed press the BACK or KNEE switch in the Siderail Control Lockouts module The padlock symbol on the control panel will be lighted when that function is locked out 3 To lock out the bed up down motion on the siderails press the Up Down s...
Страница 17: ...lenburg reverse Tren delenburg the Fowler will lower to flat the Knee will lower to flat and the litter will lower to full down 2 Press to activate the Cardiac Chair function The Knee will raise the Fowler will raise or lower to approxi mately 52 and the bed will tilt to approximately 12 reverse Trendelenburg foot end down or 14 if the bed has the enhanced height option Release the button to stop ...
Страница 18: ...d the DISARM keys and do not touch the bed until the ARMED light stops flashing 3 Put the patient on the bed and push the ARM key to activate the Bed Exit function The ARMED light will come on 4 To deactivate Bed Exit push DISARM The ARMED and BED EXIT ON lights will turn off WARNING The Bed Exit System is intended only to aid in the detection of a patient exiting the bed It is NOT intended to rep...
Страница 19: ...em 1 Before putting the patient on the bed press and hold the ARM DISARM and the SELECT ZONE keys together for 5 seconds The top light will begin to flash 2 Release the ARM DISARM and the SELECT ZONE keys and do not touch the bed until the top light stops flashing 3 Put the patient on the bed and push and release the ARM DISARM key top light will come on 4 The Bed Exit system with Zone Control wil...
Страница 20: ...he bed measures the location of the patient s center of gravity If the patient s center of gravity moves from the original location more than 6 5 inches to either side or 13 inches toward the head or foot an alarm will sound The third zone bottom indicator light is the most restrictive zone When the zone is selected the bed mea sures the location of the patient s center of gravity If the patient s...
Страница 21: ...ction may be affected by siderail caster interference With the litter fully lowered or lowered in Re verse Trendelenburg the siderails tucked under the litter in the storage position and the casters turned there is the potential for interference between the siderail and the caster Raise the siderails when lowering the litter to the full down position to prevent the interference from causing the sc...
Страница 22: ...ating the Scale System and Displaying Patient Weight page 22 3 Adding or Removing Items During a Patient s Stay page 23 4 Displaying Trendelenburg or Fowler Angle page 24 5 Converting the Patient s Weight page 24 6 Displaying the Weight Log page 25 7 Viewing Patient Weight In Gain Loss Mode page 26 8 Changing the Numerical Value Of Displayed Weight page 27 ...
Страница 23: ...TO ZERO Release The scale monitor will now read DO NOT TOUCH BED 0 0 LB The bed is now ready for the patient NOTE If there is a problem with a load cell or another component of the scale system the system will try to zero for 30 seconds and the scale monitor will read UNABLE TO ZERO TRY AGAIN If the problem continues after 3 attempts at zeroing the scale system will lock and the scale monitor will...
Страница 24: ...START Release The scale monitor will read DO NOT TOUCH BED ADD REMOVE EQUIP Add or remove the equipment and press The scale monitor will read RELEASE TO FIN Release The scale monitor will read DO NOT TOUCH BED XXX X LB The weight displayed will be that of the patient only If the CHANGE EQUIPMENT function is started but not finished after approximately 45 seconds the monitor will read HIT CH EQ TO ...
Страница 25: ...ve the scale monitor will read FOWLER ANGLE X defualt setting or TREND ANGLE X CONVERTING THE PATIENT S WEIGHT Press and release The scale monitor will read WEIGHT NOW KGS XXX X KG Repeat the procedure to return to pounds The display will read WEIGHT NOW LBS XXX X LB If the unit of measurement has been locked the display will read UNITS ARE LOCKED A service technician must be called to unlock the ...
Страница 26: ...cale monitor will read SHOW WEIGHT LOG PUSH ENTER Press The scale monitor will read WEIGHT LOG Press or to scroll through the weight log The scale monitor will read 1 0 0 LB A weight reading is logged each time the button is pressed and the bed is in the scale mode for at least 15 seconds The first weight reading displayed 1 is the most recent If the change in the patient s weight since the last r...
Страница 27: ...nter the Menu Mode The scale monitor will read FOR OPTIONS Press or to scroll through the menu options The scale monitor will read START GAIN LOSS PUSH ENTER Press The scale monitor will read DO NOT TOUCH BED XXX X LB G XX X To exit Gain Loss mode press to enter the Menu Mode The scale monitor will read FOR OPTIONS Press to to scroll through the menu options The scale monitor will read QUIT GAIN L...
Страница 28: ...read FOR OPTIONS Press or to scroll through the menu options The scale monitor will read CHNG PTNT WGT PUSH ENTER Press The scale monitor will read HOLD TO INC TO DEC Press to decrease the displayed weight or to increase the displayed weight Once the desired weight is displayed press The scale monitor will read XXX X LB NOTE If one of the load cells is malfunctioning or overloaded the scale monito...
Страница 29: ...Optional Pendant Operation 28 ...
Страница 30: ... an additional 5 seconds after the first stop The litter will raise an additional 2 inches 2 Press and hold to lower the litter 3 Press to raise the Knee section 4 Press to lower the Knee section 5 Press to raise the Back section 6 Press to lower the Back section 7 The Plug Bed In To Charge LED will be illuminated while the battery power switch is on if the battery level is low Plug the bed power ...
Страница 31: ...tially discharged but still ade quate for operation and only one will flash if the batteries are near full discharge The Plug Bed In To Charge LED on the Head End Control Panel see page 29 will remain illuminated while the battery power switch is on if the battery level is low Plug the power cord into a wall socket to charge the batteries 3 After one hour on battery power the unit will enter power...
Страница 32: ...il it locks into place at its fully raised position 3 Rotate the IV hangers B to desired position and hang IV bags 4 To lower the IV pole turn the latch C clockwise until section A lowers CAUTION The weight of the IV bags should not exceed 40 pounds To use the Removable IV pole 1 Install the pole at any of the four receptacles on the bed top located on all four corners of the bed 2 To raise the he...
Страница 33: ... is flashing the Nurse Call battery needs to be replaced The battery is located on the patient s left side at the head end of the bed No tools are required to replace the battery Unplug the bed power cord from the wall socket and replace the battery Properly dispose of the old battery in accordance with local regulations MAIN BED POWER CIRCUIT BREAKER In the event of a loss of bed functions unplug...
Страница 34: ... V pole working properly Foley bag hooks intact Chart rack intact and working properly CPR board not cracked or damaged and stores properly No cracks or splits in head and foot boards All functions on footboard working properly including LED s No rips or cracks in mattress cover Scale and Bed Exit system calibrated properly optional equipment Motion interrupt switches working properly Power cord n...
Страница 35: ... SOME CLEANING PRODUCTS ARE CORROSIVE IN NATURE AND MAY CAUSE DAMAGE TO THE PRODUCT IF USED IMPROPERLY If the products described above are used to clean Stryker patient care equipment measures must be taken to insure the beds are wiped with clean water and thoroughly dried fol lowing cleaning Failure to properly rinse and dry the beds will leave a corrosive residue on the surface of the bed possib...
Страница 36: ... in the maintenance manual for each device This statement constitutes Stryker s entire warranty with respect to the aforesaid equipment STRYKER MAKES NO OTHER WARRANTY OR REPRESENTATION EITHER EXPRESSED OR IMPLIED EXCEPT AS SET FORTH HEREIN THERE IS NO WARRANTY OF MERCHANTABILITY AND THERE ARE NO WARRANTIES OF FITNESS FOR ANY PARTICULAR PURPOSE IN NO EVENT SHALL STRYKER BE LIABLE HEREUNDER FOR INC...
Страница 37: ... without approval from the Stryker Customer Service Department An autho rization number will be provided which must be printed on the returned merchandise Stryker reserves the right to charge shipping and restocking fees on returned items SPECIAL MODIFIED OR DISCONTINUED ITEMS NOT SUBJECT TO RETURN Damaged Merchandise ICC Regulations require that claims for damaged merchandise must be made with th...
Страница 38: ...09 015 REV B European Representative Stryker EMEA RA QA Director Stryker France ZAC Satolas Green Pusignan Av De Satolas Green 69881 MEYZIEU Cedex France 3800 E Centre Ave Portage MI 49002 800 327 0770 www stryker com ...