Error - E01
Load detected does not match the
calibration procedure step. This error
may occur during the individual load
cell calibration.
•
Verify the calibration weight is in
the proper corner for the calibration
step.
•
Verify the load cell cables are
connected to the control board.
Error - E02
Load cell is installed upside down.
This error may occur during the
individual load cell calibration.
•
Verify the load cell is installed. The
error will be where the weight is
currently located on the litter. See
Error - E03
At least one load cell gives an
impossible reading. This error may
occur during normal scale system
operation.
•
Verify the load cell cable is not cut
or severed.
•
Verify the load cell cables are
connected to the control board.
•
Contact Technical Support if the
problem persists.
Error - E05
Weight on the scale exceeds the
maximum capacity. This error may
occur during normal scale system
operation.
•
Do not use the scale system when
the patient weight exceeds 100 lb
(54.4 kg).
Error - E10
Angle sensor not detected. This error
may occur during normal scale system
operation.
•
Verify the angle sensor connection
to the control board.
•
Verify the cable is not cut or
severed.
•
Contact Technical Support if the
problem persists.
Error - E11
Measured angle values are outside
target range. This error may occur
when calibrating the scale in relation to
the litter angle.
•
Make sure that the position of the
weight corresponds to the position
detailed in the procedure.
•
Verify the angle sensor connection
to the control board.
•
Restart the calibration process.
•
Contact Technical Support if the
problem persists.
Error - E12
Weight measurement taken does not
correspond to the current angle
reading. This error may occur during
normal scale system operation.
•
Verify the angle sensor connection
to the control board.
•
Recalibrate the scale system.
Error - E13
Angle reading has the wrong sign.
Trendelenburg is positive and reverse
Trendelenburg is negative. This error
may occur during the calibration of the
scale system.
•
Verify the angle sensor is installed
in the proper location.
•
Make sure that the angle sensor is
installed in the proper location.