Operation
Starting Manual mode (Continued)
Notes
•
The controller determines
Warming
or
Cooling
therapy based on the selected water target temperature and the
current water temperature.
•
In Manual mode, only the water temperature is controlled.
•
A temperature probe is not required when operating in Manual mode.
•
After the water target temperature is achieved, water temperature is controlled /-0.3° C.
Starting Monitor mode
In Monitor mode, no therapy will be delivered, only a display of the current patient temperature.
To start Monitor mode:
1.
Connect the reusable adapter cable to port A or port B on the controller. Make sure that the probe is fully seated.
2.
Apply the sensing end of patient probe to patient based on hospital protocol. Secure the patient probe to reduce the
risk of accidental dislodgment.
3.
Tap the Monitor button.
4.
Connect the patient temperature probe to the end of the reusable adapter cable. See
.
Note:
If the product senses a patient probe temperature below 36.0° C (96.8° F) or above 38.0° C (104° F), the
normothermia alarm displays and an audible alarm sounds.
5.
Tap Confirm. The screen will display the current patient temperature.
Switching modes
Tap Edit and select a different therapy mode.
Pausing and resuming therapy
To pause therapy, press and hold the Pause Therapy button for two seconds.
To resume therapy, press and hold the Pause Therapy button for two seconds.
Displaying the data storage
The system gathers data at five second intervals and is limited to 90 minutes of storage. The graphical display defaults to
show data for all four variables in the Manual and Automatic modes.