A
C
B
D
E
F
Fiig
gu
urre
e 1
14
4 –– H
Ho
oo
okk a
asssse
em
mb
bllyy
5. Using a stubby #2 Phillips screwdriver, remove the screw (B) that secures the hook assembly to the back pump
housing. Save the screw.
6. Using a T20 Torx driver, remove the six screws that secure the back pump housing to the front pump housing. Save the
screws.
7. Rotate the pump assembly to the right, so the front of the pump is on your left.
8. Separate the front and back pump housing. Place the back of the pump down on the work surface.
9. Unplug the power input cable and the power output cable that come from the power supply.
10. Using one hand to hold the fish paper and a stubby #2 Phillips screwdriver, remove the four screws and washers that
secure the power supply to the mainframe. Save the screws and washers.
11. Remove and save the power supply and fish paper.
12. Using a stubby #2 Phillips screwdriver, remove the four screws that secure the pump frame to the mainframe and set
aside. Save the screws.
13. Using a stubby #2 Phillips screwdriver, remove the six screws (E) that secure the hook assembly (D) to the back pump
housing. Save the screws.
14. Remove and discard the hook assembly.
15. Reverse steps to reinstall.
16. Plug the pump cord into an accessible outlet.
17. Run the pump diagnostic test. All tests must pass.
18. Verify proper operation before you return the product to service.
F
Frro
on
ntt p
pu
um
mp
p h
ho
ou
ussiin
ng
g rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
T20 Torx driver
•
#2 Phillips screwdriver
•
Stubby #2 Phillips screwdriver
•
ESD system
2941-009-002 Rev B.0
27
EN