Reprocessing USA and Canada
Instruction For Use • C-MAC
®
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019EN-US •
V4.3
35
3. Start the sterilization using the following parameters:
Conditioning Parameters:
Temperature: 55°C (131°F)
Humidity: 70% relative humidity
Conditioning dwell time: 30 minutes
Sterilization Parameters:
Sterilant: 100% ethylene oxide
Temperature: 55°C (131°F)
Gas concentration: 735±30 mg/l
Exposure Time: 180 minutes
Aeration Parameters
Time: At least 12 hours
Temperature: 55°C (131°F)
8.9.4 Chemical low-temperature sterilization with peracetic acid –
STERIS
®
System 1E
®
CAUTION
Processing in the STERIS System 1E
®
must occur immediately prior to use, since wet
devices cannot maintain sterility.
For sterilization with Steris
®
System 1E
®
, country-specific laws and regulations must be
observed.
Use protective cap 8403 YZ for sterilization with Steris
®
System 1E
®
(see figure).
Place the video laryngoscope in the appropriate tray.
Place the tray correctly into the SS1E and start the system.