Reprocessing
Instruction For Use • C-MAC
®
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019EN-US •
V4.3
28
Machine cleaning/chemical-thermal disinfection
A maximum temperature of 65°C must not be exceeded for this product. The relevant
national regulations and the chemical manufacturer’s instructions must be taken into
account when using this method.
7.8 Assembly, inspection and care
1. Check the product visually for purity, completeness, damage and dryness.
2. In case of soiling or residue, post-clean the product manually and carry out a complete
cleaning and disinfection process.
3. Sort out damaged and corroded products.
4. Assemble disassembled products.
5. Perform a functional check.
6. For care of the product, refer to the articles in the "Care, Sterilization and Storage
Technology" catalog.
7.9 Packaging
Only use standardized and approved packaging materials and packaging systems (EN 868
Part 2 – 10, EN ISO 11607 Part 1 + 2, DIN 58953).
7.10 Sterilization
The procedures as well as the process-relevant parameters for the individually validated
methods are described in the manual "Cleaning, Disinfection, Care and Sterilization of KARL
STORZ Instruments" (Art. no. 96216003D). The method must be selected taking into account
the respective applicable national requirements and in consultation with the device and
product manufacturers.
The following sterilization methods have been validated and approved for this product by
KARL STORZ:
7.10.1 Hydrogen peroxide (H
₂
O
₂
) sterilization - ASP STERRAD
®
WARNING
Risk of infection!
The use of STERRAD
®
processes on lubricated, oiled or pre-assembled products does not
lead to sufficient sterilization.
Sterilize only products that are not oiled and not greased.
Sterilize products in disassembled state.
WARNING
Different STERRAD® sterilization systems
Please note that materials and lumen dimensions determine what products can be sterilized
using the various STERRAD
®
sterilization systems.
ATTENTION
Follow the instructions manual for the sterilizer!
The "STERRAD
®
Sterility Guide" can be used to verify whether the relevant medical device
can be sterilized using the different STERRAD
®
devices.
The following STERRAD
®
sterilization methods have been validated and approved for this
medical device by KARL STORZ: