Electromagnetic compatibility
Instructions for use • DRILLCUT-X II/DRILLCUT-X II-35 • WWK497_EN_V1.0_03-2020_IFU_CE-MDR
34
11.5 Table 3 – Test levels for radiated and conducted
immunity tests
Guidelines and manufacturer’s declaration – electromagnetic immunity
The product is intended for use in the electromagnetic environment specified below. The user
of the product should make sure that it is used in such an environment.
Interference immunity
tests
EN/IEC 60601 test
level
Compliance
level
Electromagnetic envi-
ronment – guidelines
Conducted HF distur-
bances acc. to IEC
61000-4-6
3 V
rms
150 kHz to 80 MHz
3 V
rms
Portable and mobile HF
communications equip-
ment should be used no
closer to any part of the
product, including cables,
than the recommended
separation distance calcu-
lated from the equation
applicable to the fre-
quency of the transmitter.
Recommended separation
distances:
d = 1.2 √P
Where P is the rated
power of the transmitter in
watts [W] according to the
information provided by
the transmitter manufac-
turer and d is the recom-
mended separation dis-
tance in meters [m].
Field strengths from fixed
HF transmitters as deter-
mined by an electromag-
netic site survey
a
should
be less than the compli-
ance level in each fre-
quency range
b
.
d = 1.2 √P
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
Interferences may occur in
the vicinity of equipment
marked with the following
symbol:
Radiated HF distur-
bances acc. to IEC
61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m