ASTOPAD
®
Instructions for Use
25
7.2 On mode
Control panel
Action
Press the “Standby” button
to switch the device from
Standby mode
to
On mode
.
Device
response
•
The Standby LED
turns off.
•
The Start screen
is displayed for 1 sec.
•
The device performs a self-test. The display illuminates, the
LEDs
flash and the acoustic alarm signal sounds to
indicate that the control unit is functioning correctly.
•
When an applied part is connected, the display
shows the current actual temperature.
•
If no applied part is connected to one of the two outputs
A/B, the corresponding display is turned off.
•
Display of the saved initial set temperature.
•
For device with a built-in battery the symbol for battery sta-
tus
is displayed.
•
The self-test remains active until the heating process is
started at one of the two outputs.
•
By pressing the Standby button and simultaneously holding
the Stop Channel B button, the Start screen is displayed for
as long as the Stop button is pressed.
•
The software status can be read at the bottom right corner
of the Start screen.
9
9
12
6
10
1
2
8