4-2
920011-649
Operator Manual
Processing Instructions
4.1.2 Optional Automated
Leak Test Assembly
The automated leak test feature offered on the Reliance
™
EPS
Processor is a means to inspect endoscopes for leaks prior to
processing. The automated leak test feature allows the operator to
verify damage did not occur during the cleaning process after the
required manual leak test.
The leak test feature is also used to confirm no endoscopes are
connected during Reliance decontamination cycles (D-SHORT and
D-LONG cycles). An alarm and abort of the cycle is initiated at the
beginning of the decontamination cycle if the Reliance EPS
Processor detects an endoscope attached to the processor.
After the automated leak test is completed, a minimum pressure is
maintained in the connected endoscopes during the cycle.
In the event that the automated leak test fails, an alarm is triggered
(Fault 52 or Fault 53) and the cycle is aborted. No pressure is
maintained in the connected endoscopes.
Refer to
, for more details about the automated
leak test status (leak test passed or failed), and to
S
ECTION
7,
T
ROUBLESHOOTING
, for more details on actions user is required to
perform in case a failing leak test occurs (Fault 52 or Fault 53
triggered).
920-013-571
AUXILIARY WATER
INLET
Biopsy
Port
Air/Water
Valve Port
Air/Pipe
Air/Water
Port
Auxiliary
Water Inlet
Suction Barb
INSIDE PRESSURIZED
CONTROL HANDLE BOOT
OUTSIDE PRESSURIZED
CONTROL HANDLE BOOT
Suction
Valve Port
Figure 4-1. Solution Flow Schematic
CAUTION – POSSIBLE EQUIPMENT
DAMAGE:
Remove
any
automated leak test hoses
present
in
endoscope
processing support before
performing a D-SHORT or a D-
LONG cycle. Performing a
decontamination
cycle
(D-
SHORT and D-LONG) with
automated leak test hose
assemblies
present
may
cause water to enter hoses.
Inspect
leak
test
hose
assemblies for presence of
water prior to processing a
cycle. Water in tubing may
cause damage to endoscopes.
Содержание P920011-649
Страница 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Страница 2: ......
Страница 13: ...xi 920011 649 Operator Manual Table of Contents ...
Страница 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Страница 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Страница 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 167: ......
Страница 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...