The tank is refilled from hot and cold inlets. Detergent is dispensed and mixed with
tank water. When the tank is full, the fluids are heated and ultrasonics are used to
remove any dissolved gases. Washing alternates between ultrasonic irrigation and
lumen flushing and spraying. At the end of washing, the tank is emptied.
Rinsing removes chemical and any remaining soil from instruments. The tank is
emptied and then refilled for thermal disinfection. The water is heated to +91 °C
(+196 °F) and this temperature is maintained for 3 minutes. The tank is emptied
and cooled.
This wash program is locked and can only be adjusted by a manufacturer-approved
service technician.
12.4
P2: da Vinci Wash Cycle Program
The da Vinci Wash Cycle Program is validated to clean, with certainty, up to 10 Si/S
robotic instruments, or up to 6 Xi/X robotic instruments.
Instruments go through a pre-wash to dislodge visible soil and blood particles. The
use of cold water during this stage avoids blood clotting or the formation of protein
prions. At the end of pre-wash, the tank is emptied.
The tank is refilled from hot and cold inlets. Detergent is dispensed and mixed with
tank water. When the tank is full, the fluids are heated to and ultrasonics are used
remove any dissolved gases. Washing alternates between ultrasonic irrigation and
lumen flushing and spraying. At the end of washing, the tank is emptied.
Rinsing removes chemical and any remaining soil from instruments. The tank is
emptied after rinsing. The final stage is cold final rinse. After the wash program
completes the instruments may require further processing.
This wash program is locked and can only be adjusted by a manufacturer-approved
service technician.
12.5
P3: Standard Instruments Program
This wash program is editable using the supervisor menu. The factory settings are
shown in
Wash Programs Default Parameters
on the previous page but may not
match your equipment if the program parameters have been adjusted after
installation.
12.6
P4, P5: Open Programs
Two wash programs are populated with default
Open
wash program parameters .
These wash program parameters can adjusted within the supervisor menu to
create customer- specific wash program requirements: see
128
WASH PROGRAMS
InnoWave PCF Sonic Irrigator
Operator Manual
Содержание InnoWave PCF Sonic Irrigator
Страница 1: ...InnoWave PCF Sonic Irrigator Operator Manual ENGLISH ...
Страница 2: ......
Страница 20: ...12 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 86: ...78 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 98: ...90 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 110: ...102 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 120: ...112 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 138: ...130 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 156: ...148 This page intentionally blank InnoWave PCF Sonic Irrigator Operator Manual ...
Страница 160: ...152 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Страница 162: ...154 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...
Страница 164: ...156 APPENDICES InnoWave PCF Sonic Irrigator Operator Manual This page intentionally blank ...