4-6
122998-741
Operator Manual
Operating Instructions
4.3.3 Alkaline Wash Phase
Hot water from building supply fills sump. If temperature is guaran-
teed, actual sump temperature alternates with guaranteed set tem-
perature:
CYCLE 10
(G=)137.4 F
ALK. WASH FILL
2:00
If conductivity option is present:
1.
Water is heated while recirculating. If temperature is guaranteed
actual sump temperature alternates with guaranteed set
temperature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
CYCLE 10
(G=)141.5 F
RECIRC/HEATING
5:00
2.
Alkaline detergent is injected into sump. If temperature is guar-
anteed, actual sump temperature alternates with guaranteed set
temperature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
CYCLE 10
(G=)141.5 F
INJECTING ALK.
5:00
3.
Detergent solution is recirculated through oscillating jet system
for programmed amount of time. If temperature is guaranteed,
actual sump temperature alternates with guaranteed set temper-
ature.
NOTE: Control alternates between these display screens every
four seconds during recirculation. Displayed time will only count
down when solution temperature is equal to or greater than set
point.
CYCLE 10
(G=)140.0 F
ALKALINE WASH
5:00
4.
Printer records time and solution temperature in chamber sump
at beginning and end of recirculation.
ALK. WASH10:21:55A
141.5
10:26:55A
140.2
*************************************
4.3.4 Time-Based pH
Neutralization (Option)
Alkaline neutralizer is injected during programmed amount of time. If
temperature is guaranteed, actual sump temperature alternates with
guaranteed set temperature.
CYCLE 10
(G=)167.8 F
ALK NEUTRALIZE
2:00
WARNING – BURN HAZARD:
Water discharge may be
extremely hot.