6-32
P129394-020
Operator Manual
Sterilizer Operation
6.16 Cycle Graphs
These cycle graphs provide a visual representation of AMSCO 400
Series Steam Sterilizer cycles and their phases.
Figure 6-16. Cycle Graph – 270°F or 275°F Prevacuum and Air Removal (Bowie-Dick) Cycles
Indicates Key Cycle
Transition Points That
Are Printed During Cycle
Press
ur
e
Time
*
Pulses
*
NOTE:
270°F (132°C) prevac cycle runs four conditioning vacuum pulses. 275°F (135°C) prevac runs three
conditioning vacuum pulses.
Purge
Charge
Fast
Exh
Dry
A
ir
B
re
ak
R
et
ra
ct
S
ea
ls
C
o
m
p
le
te
Exhaust
Sterilize
Conditioning
A
c
ti
v
a
te
S
e
a
l(
s
)
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...