9-15
Troubleshooting
Operator Manual
P129394-020
Alarm
Description
Screen with Operator Instructions
9.3.2 Pressure In
Chamber
Occurs if 2 psig
pressure is sensed in
the chamber. is greater
than 2 psig, but less
than 50 psig.
NOTE: Door can be
open or closed.
Screen with
Service
Instructions
Alarm
Description
Screen with Operator Instructions
9.3.3 Atmospheric
Pressure Alarm
Occurs if atmospheric
pressure is greater than
1 psi (or 2 psig, if door
is open) from the
calibrated atmospheric
pressure.
STATUS . . ALARM!
221
PRESSURE IN CHAMBER
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY EXHAUST CHAMBER TO
ATMOSPHERIC PRESSURE
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
222
PRESSURE IN CHAMBER
-> 2 PSIG PRESSURE SENSED IN CHAMBER
WHEN NOT IN CYCLE
CAUSES AND CORRECTIONS:
1. SOLENOID VALVE MALFUNCTION
-> REPAIR S02
2. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
3. TRANSDUCER, CP, MALFUNCTION
-> REPAIR
-> RECALIBRATE
4. MAIN CONTROL FAILURE
-> CHECK CONTROL BOARD POWER STATUS LEDs
-> REPLACE CONTROL BOARD
-> RECALIBRATE
PAPER
FEED
EXIT
STATUS . . ALARM!
269
ATMOSPHERIC PRESSURE ALARM
CHAMBER: 00.0 F 0.0 psig
ALTITUDE SETTING: 0 TO 1000 FEET
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...