![Steris AMSCO 400 Series Скачать руководство пользователя страница 152](http://html.mh-extra.com/html/steris/amsco-400-series/amsco-400-series_operators-manual_1374315152.webp)
9-10
P129394-020
Operator Manual
Troubleshooting
Alarm
Description
Screen with Operator Instructions
9.2.12 Too Long In Air
Break
Occurs if chamber does
not air break the
vacuum to 2 inHg within
the allotted time.
Screen with
Service
Instructions
Alarm
Description
Screen with Operator Instructions
9.2.13 Too Long To
Unseal Door
Occurs if door seal
pressure does not drop
below 5 psig within
allotted time.
STATUS . . ALARM!
225
TOO LONG IN AIR BREAK
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
AUTOMATICALLY TRY TO COMPLETE CYCLE
•
EXTEND EXHAUST TIME
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. IF ALARM RECURS, CALL SERVICE
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
STATUS . . SERVICE INFORMATION:
226
TOO LONG IN AIR BREAK
-> CHAMBER DID NOT AIR BREAK
VACUUM TO 1 PSIG WITHIN ALLOTTED TIME
CAUSES AND CORRECTIONS:
1. AIR INLET FILTER PLUGGED
-> REPAIR
2. SOLENOID VALVE MALFUNCTION
-> REPAIR S01
3. CONTROL OUT OF CALIBRATION
-> RECALIBRATE (CONTACT QUALIFIED
SERVICE PERSON)
EXIT
PAPER
FEED
STATUS . . ALARM!
229
TOO LONG TO UNSEAL DOOR
CHAMBER: 00.0 F 0.0 psig
STERILIZER WILL:
•
REMAIN IN ALARM CONDITION UNTIL
DOOR IS UNSEALED
OPERAT0R INSTRUCTIONS:
1. SILENCE ALARM
2. CALL SERVICE
3. IF LOAD MUST BE REMOVED, REFER TO
EMERGENCY DOOR OPERATION PROCEDURE
IN OPERATING MANUAL
SERVICE
HELP
STATUS
SILENCE
ALARM
PAPER
FEED
Содержание AMSCO 400 Series
Страница 2: ......
Страница 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Страница 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Страница 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Страница 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Страница 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Страница 80: ...6 20 P129394 020 Operator Manual Sterilizer Operation Figure 6 10 Typical Printout of a Liquid Cycle ...
Страница 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Страница 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...