5-9
Routine Maintenance
Operator Manual
150832-639
spare parts. If batteries are procured and stored, they should
be kept covered and in a cool, dry area. Stored batteries must
be recharged every six months to minimize deterioration. Use
a charging current commensurate with the battery amp-hour
size. Charge to a floating charge voltage equivalent to 13.6-
13.8 volts.
5.8 ALS Check
The STERIS 4085 General Surgical Table is equipped with the Auto
Limit Sensor
™
system. This system provides instant user feedback on
any conflicting tabletop articulations. An Auto Limit Sensors (ALS)
Warning (refer to Fig. 4-4, Item 2) on the hand control indicates the
tabletop has been ordered into a configuration or has moved into a
position that may cause tabletop damage or patient injury. To
prevent a tabletop sectional conflict, or collision of tabletop with the
table or floor, ALS System immediately stops the movement in
process and the ALS LED lights. The control system alerts the user
to an alternative movement choice in the LCD Display (refer to
Figure 4-4, Item 1).
NOTE: The ALS System suggested alternative movement is not the
only movement available. Users should exercise professional
judgement in determining the most desirable movement.
Test the Auto Limit Sensor system as follows:
1.
Review and follow instructions presented in
,
and prepare the STERIS 4085 General
Surgical Table for operation.
2.
Raise tabletop to highest position possible.
3.
Use
SLIDE HEAD
and move tabletop to head end limit.
4.
Press
LEG DOWN
button until ALS Warning lights and tabletop
motion stops. If system does not engage, stop tabletop move-
ment before collision and call STERIS.
Table 5-3. Replacement Parts
DESCRIPTION
PART NUMBER
RECOMMENDED
SPARES
Hand Control Assembly
150832-632
1
Power Cord
093909-354
1
Fuse (Main), 4.0 Amp
150832-335
1
Fuse (Set)
342213555
1
Pad, Head Section
542200239
1
Pad, Back/Seat Section
542200240
1
Pad, Leg Section
542200241
1
WARNING – PERSONAL
INJURY AND/OR EQUIPMENT
DAMAGE HAZARD: The Auto
Limit Sensor (ALS) System
suggested alternative tabletop
movement does not take into
account the type of surgery
that may be in process or the
environment (e.g., other
equipment close by) and
might not be compatible with
the surgical procedure. To
avoid an action that could
result in death or serious
injury to the patient and the
operating room staff, and/or
serious equipment damage,
completely evaluate the
feasibility of the suggestion.
Содержание 4085
Страница 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Страница 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Страница 61: ...MAINTENANCE MANUAL Maintenance Instructions STERIS 4085 General Surgical Table 07 01 10 P764333 766 ...
Страница 131: ......
Страница 161: ...32 764333 767 Figure 5 4085 Surgical Table Hydraulics 1 of 2 4 3 11 12 8 7 24 18 19 22 23 15 14 9 10 6 5 See 2 of 2 1 2 ...
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ......
Страница 177: ......