3.3. Odyssey Quad HD System Verification
No.
Test Description
Pass/Fail
Initials
1
Verify Odyssey Quad HD Cabinet labeling. Verify the Odyssey Quad HD
Cabinet label is affixed to the upper right front of the Odyssey Cabinet.
•
Record the Odyssey P/N from the label: ___________________
Y / N
N/A
_____
2
Verify Odyssey System labeling. Verify the Odyssey Quad HD System label
(HDW-0306) is affixed to:
•
The upper right of the Odyssey Quad HD cabinet
•
On the TPI’s next to the part number label
•
On the Procedure Room Hub right side where the cables enter
P F
P F
P F
_____
_____
_____
3
Verify Navigant Software Versions.
•
Start the Odyssey system and record the Navigant software version that is
located on the main menu window:
_______________________________________________
•
For MDM Software Version. End the Navigant software (Ctrl>Alt>Del, at
task manager select the Applications tab and locate Navigant.
Workstation. Select End Task). Select the Firefox icon, and then select
the MDM tab. Select about. Record the MDM software version here:
_______________________________________________
•
For systems that include an OBUI table side controller, record the
firmware version that appears on boot up (For Niobe Systems with 3.2
S/W and above and Std Labs).
_______________________________________________
•
Verify these software versions are the latest released versions. If not,
provide explanation here:
_______________________________________________
N/A
P F
_____
_____