Upon power-up of instrument, please allow 5 minutes for instrument warm-up.

Upon thaw completion, the vial is raised 

approximately 1 cm (0.4 in).

Remove vial immediately and for a few 

seconds, manually agitate the vial gently 

from side to side to equilibrate the liquid 

temperature until no ice can be seen within 

the vial. Begin downstream protocol. 

Countdown of the last 10 seconds of the 

thawing process begins.

Beginning of solid to liquid phase change is 

detected by instrument.

Instrument is warmed up and in “ready” state. 

Insert a clean, tightly capped frozen 

1.8 - 2.0 mL cryogenic vial (below -70°C) 

containing up to 1.8 mL volume into the 

opening on the top of the instrument.

Push the top of the vial down gently until 

the instrument emits a deep dual tone. 

The vial will now remain in a depressed 

position.

 

 

ThawSTAR

®

 

CFT2 

Quick Start Guide