9.1.14. Display Resolution
The number of decimal places that are used to display the detected gas
levels can be changed via this menu option.
•
From the menu system select menu option:
Display
and press
ENTER.
Display
Edit
•
Using the
buttons to move the decimal point to the
desired position, e.g.
Display
Edit
.
•
Press ENTER to accept the new alarm level,
will be
displayed
Note: Pressing MENU instead of ENTER leaves the display
setting unchanged.
•
Press the MENU button to exit the display option menu. On exit
the display will show if no changes have been
made or if changes have been made.
Note that when the sensor type is first selected, the optimum number of
decimal places for a given sensor type and range is automatically selected.
Increasing the number of decimal places to which detected gas levels are
displayed does not increase the sensitivity or accuracy of the instrument and
in some cases may degrade the instruments perceived stability.