background image

D2432-01-09 CN5171 SEM203TC  User Guide

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

1.0 

DESCRIPTION 

  

The  SEM203TC  is  an  in-head  (4  to  20)  mA  transmitter  that  connects  to  a  standard 
thermocouple  sensor and  converts  the  sensors  temperature  to a  linear temperature  (4  to 20) 
mA signal.  
The transmitter sensor type and range can be requested at the time of order, but if desired the 
user can re-configure the transmitter parameters by use of a single push button and the range  
“R” , menu “M” LEDS.  Two methods of configuration are available, the first “

USER RANGING

” 

acts  only  on  the  transmitter  range,  similar  to  the  previous  SEM203TC  design.  The  other 
method “

ADVANCED USER CONFIGURATION

 “ offers full configuration, This level is entered 

by holding down the push button on power up. The advanced level has the following menus:- 
 

 

Menu 1 - Selection of input type , seven popular thermocouples or mV input. 

 

Menu 2 - Select either user push button set range or one of seven fixed ranges. 

 

Menu 3 - Select either up or down scale output on sensor burnout. 

 

Menu 4 - User trim allow trim of output current at high and low range 

 

Menu 5 - Reset factory default. 

 
 

The addition of fixed ranges to this product allows the used to re-range the product without the 
need for specialist equipment. The SEM203TC input is isolated. 

 
2.0 

RECEIVING AND UNPACKING 

 

Please inspect the packaging and instrument thoroughly for any signs of transit damage. If the 
instrument has been damaged, please notify your supplier immediately. 
 

3.0 

SPECIFICATION @ 20 °C 

 
INPUT 

 

Sensor 

Range (

°

C) 

Accuracy 

 

-200 to 1370 

± 0.1% of F.S.  ± 0.5 °C  
(plus any sensor error) 

-100 to 1200 

 

± 0.1% of F.S.  ± 0.5 °C  
(plus any sensor error) 

-100 to 1000 

± 0.1% of F.S.  ± 0.5 °C  
(plus any sensor error) 

-180 to 1300 

 

± 0.1% of F.S.  ± 0.5 °C  
(plus any sensor error) 

-100 to 400 

 

± 0.2% of F.S.  ± 0.5 °C  
(plus any sensor error) 

R  

-10 to 1760 

 

± 0.1% of F.S.  ± 0.5 °C  
(plus any sensor error) over range 800 to 1600 

S  

-10 to 1760 

± 0.1% of F.S.  ± 0.5 °C  
(plus any sensor error) over range 800 to 1600 

 

Range (mV) 

 
 

mV 

-10 to 70  

± 0.02% of full scale  

 

 

Isolation 

 

Tested to 250V dc

 

Sensor Burnout                 

Either up or down scale output 

Cold Junction 

Range (-40 to 85) °C; Accuracy  ± 0.5 °C  
Tracking ± 0.05 °C / °C 

Stability 

 

Offset 0.15 °C / °C  ; Span 0.1 °C / °C

 

 
OUTPUT 
 
Type 

 

Two wire (4 to 20) mA sink

 

Limits 

 

Low 3.8 mA ; high 21.5 mA

 

Accuracy 

 

(mA out / 2000) or 5 uA which ever greater

 

Loop Effect   

± 0.2 uA / V  measured @ 50 Hz 1V (peak to peak

Thermal Drift  

 ± 2 uA 

Max Load 

 

[(Vsupply – 12)/20] K

 

 
GENERAL 
 
Operating Voltage

 

(12 to 30) Volts DC

 

Update Time   

0.5 Seconds

 

Response Time 

1 Second to reach 90% of final value

 

Start up time   

From power up typically  5 Seconds

 

Filter Factor   

Adaptive

 

Ambient Temperature 

(-40 to 85) °C

  

Connection   

Screw Terminal

 

Approvals 

BS EN 61326 ; – Electrical equipment for measurement and 
control  ANNEX A ; ANNEX F 

Factory Default 

(0 to 1000) °C type K, upscale burnout (0.0 °C user trim) 

 

 

4.0 

INSTALLATION AND WIRING 

  

4.1 Mechanical 
 

The SEM203TC transmitter has been specifically designed to fit inside a DIN standard probe 
head enclosure (such as the Status SCH-4 series), which provides adequate protection  from  
moisture, dust,  corrosive atmosphere etc. All cable entries must be sealed using the correct 
size gland. Likewise any probe assembly fitted must be sealed. 
Care must be taken when locating the transmitter to ensure the working ambient temperature 
range of (-40 to 85) °C is not exceeded.  The SEM203TC enclosure has a centre hole allowing 
the sensor wired to enter screw terminals from the transmitter centre, this is applicable when 
the sensor is mounted directly below the transmitter. 

 
 
 
 
 
 
 

 
Figure 1 
 
Mounting holes : two holes 5.5 mm diameter, 33 mm centres 
Centre Hole sensor wire entry : 4mm 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
4.2 Electrical 

 
Electrical connections to the transmitter are made to the screw terminal provided on the top 
face. The correct type of thermocouple wire must be used to connect sensor, this will normally 
be provided as part of the probe assembly. The screw terminals allow for wires to enter either 
inner or outer direction. Never attempt to unscrew the spare terminal which secures the factory 
fitted cold junction sensor. The terminal is fitted with a tamperproof screw to avoid accidental 
adjustment. 
The transmitter is protected  against  reverse connection and over voltage. If no sensor (input) 
connection is made the transmitter will go into either up or down scale output current, 
depending on configuration setting. 
Figure 2  gives connection details, the output is shown connected to a 24 V supply. The load 
symbol represent any other device connected in the loop, such as Monitoring equipment , 
panel indicators and loop isolators. The load value can range from 0 ohms to the max loop load 
for given supply, refer to section 3  “Max load”  for more information. 
The transmitter conforms with EC directive BS EN 61326 : when correctly installed in a 
termination head providing at least  IP20 protection and with sensor wires less than 3 metres. 
Screened or twisted pair wires are recommended for output wires. Always ensure the (4 to 20 ) 
mA loop is grounded at one point, this would normally be at the monitoring equipment or loop 
power supply. 
In normal operation the range “R” LED acts as over-range LED.  The menu LED is always off.  

 
Figure 2

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

SEM203/TC 

PUSH BUTTON  

THERMOCOUPLE TX 

 
Designed, manufactured and supported by: 

 

Status Business Park, Gannaway Lane 

Tewkesbury Glos. G20 8FD. UK 

Tel: +44 (0)1684 296818    Fax: +44 (0)1684 293746 

Email: 

[email protected]

 

 
 
Every effort has been taken to ensure the accuracy of this specification, however we do 
not  accept  responsibility  for  damage,  injury,  loss  or  expense  resulting  from  errors  and 
omissions, and we reserve the right of amendment without notice 
 

 

 

 

STOCK CODE

  

52-315-2432-01

 

Ø43.0 mm

21.0 mm

33.0 mm

Centres

U

-

+

-

+

R

M

T/C

LOAD

Отзывы: