Installation - Swivel Motors

 

Installation - Non Swivel Motors

• Operation
• Cleaning 
• Sterilization

  5,000 rpm Motor:   Titan

®

(

StarDental

®

 

Part #262261); 

Titan 3 

For use with Extended Straight Attachment 

(ESA) 

(

StarDental

®

 

Part #265208); 

Titan 3 Fixed Back End 

(

StarDental

 

Part #262409); or

 

Titan T 

(

StarDental

 

Part #263940)

20,000 rpm Motor:   Titan 3

 (

StarDental

 Part #262262); 

Titan 3 Fixed Back En

d (

StarDental

 

Part 262410); or

 

Titan T 

(

StarDental

 Part #263941)

 

• No Lubrication Required

 

• Recommended Air Pressure: 40-45 psi (swivel motor)

 

• Air Flow: 2.1 SCFM @ 40-45 psi at Maximum Power Point (swivel motor)

 

• Rotation: Forward and Reverse

 

• 4-Line Connection (Fixed Back End)

Installation - Swivel Motors

1.  Select the motor appropriate for tubing based on power requirements. Examine  connection of the motor and 

swivel

 tubing configuration as these must match configuration of delivery 

system tubing.

Drive Air

Water

Exhaust Air

Air

4-Line Tubing

2/3-Line Tubing

Drive Air

Water

2.  Disconnect swivel from motor if necessary. Grasp the swivel in front of the bearing assembly between the 

thumb and

 index finger as shown.

3

.  Slide  tubing  coupler  until  the  end  of  the  ISO  connector  is  exposed;  hold  in  hand  as  shown

  with  fingers 

securing the tubing.

4. Align and insert swivel tubes into the ISO connector holes.
5.  While  applying  slight  pressure  to  hold

  fittings  in  the  connector  holes,  slide 

coupler  forward  and  engage 

threads. Continue threading coupler clockwise until hand tight against the swivel.

 

Swivel

Bearing Assembly

ISO Connector

Tubing 

Coupler

Tubing

6.  Using the wrench as shown, tighten the swivel to the tubing coupler. Be sure to turn tubing coupler, not swivel. 

DO NOT OVER TIGHTEN.                   

Wrench

Tubing 

Coupler

Turn Coupler While Holding     

         Swivel in Place

Note:  

This  joint  loosens over time, resulting in air leakage and  loss of power. Check  for  tightness  periodically.

7. Connect motor to swivel.

Operation

The 120° rotation speed adjustment ring sets direction and speed. To run motor forward at full speed, rotate ring towards 

the letter ‘F’ while operating the foot control. Rotate to the letter ‘R’ to yield full speed in reverse. Vary speed by rotating ring 

in either direction towards center. Note the small depression for the current setting. After setting, the motor runs at current 

speed and direction until reset. Pausing the motor is not needed to alter direction or speed using the ring.

20,000 rpm - Lube Free

Made in the U.S.A.

JXXXXX

®

Titan 3

StarDental

®

Small Depression

120° Rotation Speed 

Adjustment Ring

• Application Guide
• Troubleshooting
• Replacement Part Numbers
• Warranty

®

Titan 3

  5,000 rpm - Lube Free

Made in the U.S.A.

JXXXXX

StarDental

®

Titan 3

20,000 rpm - Lube Free

Made in the U.S.A.

JXXXXX

®

StarDental

®

5K Motor

20K Motor

Swivel

Drive Gear

Installation - Non Swivel Motors

Note:

 If delivery unit has oil/mist lubrication, detach and drain; motor needs no lubrication.

NOTICE

 To prevent damage to bearings, before installing always check drive air pressure by using air pressure 

gauge, 

StarDental

 Part #262591 (4/6-Line) between motor and tubing and obtain a reading. Adjust air pressure 

at delivery head until gauge reads 40-45 psi while operating motor in full forward postion and with foot control 

completely depressed.

1. Examine rear end connector of motor. Verify motor matches tubing connector of delivery system.
2. To connect motor to tubing, align drive air and exhaust tubes with mating ISO connector holes in coupler.
3.  While applying slight pressure to hold tubing in connector holes, slide tubing coupler forward and engage 

threads. Continue threading tubing coupler clockwise until hand tight against the motor. DO NOT OVER TIGHTEN.

Note:

  This joint  loosens  over  time,  resulting in air leakage  and  loss of power. Check  for  tightness  periodically.

®

Titan 3

  5,000 rpm - Lube Free

Made in the U.S.A.

JXXXXX

StarDental

®

Tita

n3

20,00

0 rp

m - L

ube

 Fr

ee

Mad

e in

 th

e U

.S.

A.

JXX

XXX

R

Star

Dental

In

str

um

en

ts

®

Direction & Speed

Control Ring

Drive 

Gear

Rear End Connector

5K Motor 

(

StarDental

®

 Part #262409)

Drive 

Gear

20K Motor 

(

StarDental

®

 Part #262410)

Sterilizable up to 135°C

135°C

86

6.

DT

E.

IN

FO

 (3

83

.4

63

6)

w

w

w

.D

en

ta

lE

Z.

co

m

03

44

St

ar

D

en

ta

18

16

 C

ol

on

ia

l V

ill

ag

La

ne

La

nc

as

te

r, 

PA

 17

60

1  

U

.S

.A

.

Pr

in

te

in

 th

U

.S

.A

.

©

20

19

 D

en

ta

lE

Z,

 In

c.

Em

er

go

 E

ur

op

e

Pr

in

se

ss

eg

ra

ch

t 2

0

25

14

 A

Th

H

ag

ue

Th

N

et

he

rl

an

ds

EC

RE

P

StarDental®

 Instruments

Ca

ut

ion:

 F

eder

al la

w r

es

tr

ic

ts

 th

is

 d

ev

ic

to

 

                     sale 

by

 or on the or

der of a den

tist.

 

06

22

83

  P

  D

at

of

 Is

su

 2

01

9-

05

-2

4

Titan® Lo

w

-Speed 

Handpiece Motor

Instruction Manual

N

on

-S

te

ri

le

Medical Device

Indications For Use

BEFORE OPERATING THE MOTOR, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE 

THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.

The Titan 5K and 20K low speed dental motors are used by trained dental professionals to facilitate various 

dental procedures requiring cutting, drilling, polishing and pin setting.

Note:

 If delivery unit has oil/mist lubrication, detach and drain; motor needs no lubrication.

NOTICE

 To prevent damage to bearings, before installing always check drive air pressure by using air pressure 

gauge, 

StarDental

 Part #262591 (4/6-Line) between swivel and tubing and obtain a reading. Adjust air pressure 

at delivery head until gauge reads 40-45 psi while operating motor in full forward position and with foot control 

completely depressed.

Cleaning & Sterilization

CAUTION

INSTRUCTIONS

6WDU

7LWD

U

SP

/X

EH)

UHH

0DG

HLQ

WKH

86

$

-;;

;;;

5

5

StarDental®

 In

strum

ent Solutions

Drive Air

Red Tip  

Applicator

Maintenance &Lubrication:

The 

Titan

 Motor is LubeFree and, therefore, requires no lubrication.

However, at times debris from the delivery system’s drive air tubing collects in the motor. This build-up may cause 

sluggish motor operation. To ease, place 3 drops of DentaLube 

II

StarDental

 Part #262539 into the drive air tube of 

motor, and alternately operate the motor in forward and reverse.

Motor should be routinely maintained, checked for corrosion, worn parts, etc. and should be replaced on an as needed basis.

WARNING

Do not exceed 135°C

Do not use:

Limits on Reprocessing

Repeated processing has minimal effect on these attachments. End of life is normally determined by wear and 
damage due to use.

  The handpiece motor must be cleaned and sterilized before first use and between patient use.

Star

Titan

3

20,00

0 rpm

 - L

ube F

ree

Made

 in t

he U

.S.A

.

JXX

XXX

R

R

Internal  

Surfaces

Swivel Bearing/ 

Spring Assembly

Clean

Cleaning-Manual:

Cleaning the Swivel &Mating Motor Part: To optimize operation of the swivel connector, clean all internal surfaces of the 

motor with a cotton swab soaked in isopropyl alcohol until clean. Clean area around bearing/spring assembly at rear of 

swivel using isopropyl alcohol and a toothbrush. Dry both 

using syringe air.
Sterilizing Handpiece Motor: 

After each patient, flush 

air by running handpiece motor for 20 seconds.

Remove attachment. Clean all internal surfaces at rear 

of handpiece motor with a cotton swab soaked in 

isopropyl alcohol until clean. Clean all external surfaces 

using a small toothbrush with isopropyl alcohol. Dry 

thoroughly. Place handpiece motor and swivel in 

sterilization bag – . Follow the sterilization guidelines as 

stated by the manufacturer of your sterilizer.

Cleaning-Automated:

Washer-disinfectors are not recommended.

Wrapped Instrument 

Sterilization:

Gravity Type Autoclave:                                                  Pre-Vacuum Autoclave: 

• Minimum of 15 minutes at 132° C                            • Minimum of 4 minutes at 132° C
• Minimum of 30 minutes at 121° C                                   

Do not exceed 135° C

 • dry heat sterilizer

•immersion techniques

• cold sterilizers • ultrasonic cleaners

• sterilizer bags containing disinfectants • hand soaps

• detergents

• chemical disinfectors

•surface sprays

• surgical milk

• surface wipes

Additional Information:

Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only paper or paper/plastic bags.

Storage:

Store in bag after sterilization until ready to use.

Dry the motor for 30 minutes minimum using the sterilization dry cycle and allow the motor to cool to room 
temperature in bag prior to use. If moisture is present, increase your sterilizer dry cycle appropriately.

Drying:

Containment and transportation:

It is recommended that handpieces are reprocessed as soon as is reasonably practical following use.

Preparation for cleaning:

Remove motor from swivel if applicable or tubing.

Use steam sterilization exclusively for complete sterility as outlined below.

Disinfection:

See Installation, Operating Recommendations, and/or Troubleshooting sections of this manual.

Inspection & function testing:

Single:  A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain the 
item without stressing the seals.

Packaging:

Manufacturer Contact:

See first or last panels of this manual.

It is the responsibility of the user to use only FDA compliant sterilizing equipment and accessories that are capable of 

meeting the manufacturers recommendations below.

CAUTION

NOTICE

Any serious incident occurring from the use of this device should be reported to StarDental immediately. 

If incident occurred in a EU Member State the competent authority must be notified.