Contents
Specifications
•
Specifications
•
Installation - Swivel Motors
•
Installation - Non Swivel Motors
• Operation
• Cleaning
• Sterilization
5,000 rpm Motor: Titan
®
3
(
StarDental
®
Part #262261);
Titan 3
For use with Extended Straight Attachment
(ESA)
(
StarDental
®
Part #265208);
Titan 3 Fixed Back End
(
StarDental
Part #262409); or
Titan T
(
StarDental
Part #263940)
20,000 rpm Motor: Titan 3
(
StarDental
Part #262262);
Titan 3 Fixed Back En
d (
StarDental
Part 262410); or
Titan T
(
StarDental
Part #263941)
• No Lubrication Required
• Recommended Air Pressure: 40-45 psi (swivel motor)
• Air Flow: 2.1 SCFM @ 40-45 psi at Maximum Power Point (swivel motor)
• Rotation: Forward and Reverse
• 4-Line Connection (Fixed Back End)
Installation - Swivel Motors
1. Select the motor appropriate for tubing based on power requirements. Examine connection of the motor and
swivel
tubing configuration as these must match configuration of delivery
system tubing.
Drive Air
Water
Exhaust Air
Air
4-Line Tubing
2/3-Line Tubing
Drive Air
Water
2. Disconnect swivel from motor if necessary. Grasp the swivel in front of the bearing assembly between the
thumb and
index finger as shown.
3
. Slide tubing coupler until the end of the ISO connector is exposed; hold in hand as shown
with fingers
securing the tubing.
4. Align and insert swivel tubes into the ISO connector holes.
5. While applying slight pressure to hold
fittings in the connector holes, slide
coupler forward and engage
threads. Continue threading coupler clockwise until hand tight against the swivel.
Swivel
Bearing Assembly
ISO Connector
Tubing
Coupler
Tubing
6. Using the wrench as shown, tighten the swivel to the tubing coupler. Be sure to turn tubing coupler, not swivel.
DO NOT OVER TIGHTEN.
Wrench
Tubing
Coupler
Turn Coupler While Holding
Swivel in Place
Note:
This joint loosens over time, resulting in air leakage and loss of power. Check for tightness periodically.
7. Connect motor to swivel.
Operation
The 120° rotation speed adjustment ring sets direction and speed. To run motor forward at full speed, rotate ring towards
the letter ‘F’ while operating the foot control. Rotate to the letter ‘R’ to yield full speed in reverse. Vary speed by rotating ring
in either direction towards center. Note the small depression for the current setting. After setting, the motor runs at current
speed and direction until reset. Pausing the motor is not needed to alter direction or speed using the ring.
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
Titan 3
StarDental
®
Small Depression
120° Rotation Speed
Adjustment Ring
• Application Guide
• Troubleshooting
• Replacement Part Numbers
• Warranty
®
Titan 3
5,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
StarDental
®
Titan 3
20,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
®
StarDental
®
5K Motor
20K Motor
Swivel
Drive Gear
Installation - Non Swivel Motors
Note:
If delivery unit has oil/mist lubrication, detach and drain; motor needs no lubrication.
NOTICE
To prevent damage to bearings, before installing always check drive air pressure by using air pressure
gauge,
StarDental
Part #262591 (4/6-Line) between motor and tubing and obtain a reading. Adjust air pressure
at delivery head until gauge reads 40-45 psi while operating motor in full forward postion and with foot control
completely depressed.
1. Examine rear end connector of motor. Verify motor matches tubing connector of delivery system.
2. To connect motor to tubing, align drive air and exhaust tubes with mating ISO connector holes in coupler.
3. While applying slight pressure to hold tubing in connector holes, slide tubing coupler forward and engage
threads. Continue threading tubing coupler clockwise until hand tight against the motor. DO NOT OVER TIGHTEN.
Note:
This joint loosens over time, resulting in air leakage and loss of power. Check for tightness periodically.
®
Titan 3
5,000 rpm - Lube Free
Made in the U.S.A.
JXXXXX
StarDental
®
Tita
n3
20,00
0 rp
m - L
ube
Fr
ee
Mad
e in
th
e U
.S.
A.
JXX
XXX
R
Star
Dental
In
str
um
en
ts
®
Direction & Speed
Control Ring
Drive
Gear
Rear End Connector
5K Motor
(
StarDental
®
Part #262409)
Drive
Gear
20K Motor
(
StarDental
®
Part #262410)
Sterilizable up to 135°C
135°C
86
6.
DT
E.
IN
FO
(3
83
.4
63
6)
w
w
w
.D
en
ta
lE
Z.
co
m
03
44
St
ar
D
en
ta
l®
18
16
C
ol
on
ia
l V
ill
ag
e
La
ne
La
nc
as
te
r,
PA
17
60
1
U
.S
.A
.
Pr
in
te
d
in
th
e
U
.S
.A
.
©
20
19
D
en
ta
lE
Z,
In
c.
Em
er
go
E
ur
op
e
Pr
in
se
ss
eg
ra
ch
t 2
0
25
14
A
P
Th
e
H
ag
ue
Th
e
N
et
he
rl
an
ds
EC
RE
P
StarDental®
Instruments
Ca
ut
ion:
F
eder
al la
w r
es
tr
ic
ts
th
is
d
ev
ic
e
to
sale
by
or on the or
der of a den
tist.
06
22
83
P
D
at
e
of
Is
su
e
2
01
9-
05
-2
4
Titan® Lo
w
-Speed
Handpiece Motor
Instruction Manual
N
on
-S
te
ri
le
Medical Device
Indications For Use
BEFORE OPERATING THE MOTOR, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE
THESE INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
The Titan 5K and 20K low speed dental motors are used by trained dental professionals to facilitate various
dental procedures requiring cutting, drilling, polishing and pin setting.
Note:
If delivery unit has oil/mist lubrication, detach and drain; motor needs no lubrication.
NOTICE
To prevent damage to bearings, before installing always check drive air pressure by using air pressure
gauge,
StarDental
Part #262591 (4/6-Line) between swivel and tubing and obtain a reading. Adjust air pressure
at delivery head until gauge reads 40-45 psi while operating motor in full forward position and with foot control
completely depressed.
Cleaning & Sterilization
CAUTION
INSTRUCTIONS
6WDU
7LWD
Q
U
SP
/X
EH)
UHH
0DG
HLQ
WKH
86
$
-;;
;;;
5
5
StarDental®
In
strum
ent Solutions
Drive Air
Red Tip
Applicator
Maintenance &Lubrication:
The
Titan
Motor is LubeFree and, therefore, requires no lubrication.
However, at times debris from the delivery system’s drive air tubing collects in the motor. This build-up may cause
sluggish motor operation. To ease, place 3 drops of DentaLube
II
,
StarDental
Part #262539 into the drive air tube of
motor, and alternately operate the motor in forward and reverse.
Motor should be routinely maintained, checked for corrosion, worn parts, etc. and should be replaced on an as needed basis.
WARNING
Do not exceed 135°C
Do not use:
Limits on Reprocessing
Repeated processing has minimal effect on these attachments. End of life is normally determined by wear and
damage due to use.
The handpiece motor must be cleaned and sterilized before first use and between patient use.
Star
Titan
3
20,00
0 rpm
- L
ube F
ree
Made
in t
he U
.S.A
.
JXX
XXX
R
R
Internal
Surfaces
Swivel Bearing/
Spring Assembly
Clean
Cleaning-Manual:
Cleaning the Swivel &Mating Motor Part: To optimize operation of the swivel connector, clean all internal surfaces of the
motor with a cotton swab soaked in isopropyl alcohol until clean. Clean area around bearing/spring assembly at rear of
swivel using isopropyl alcohol and a toothbrush. Dry both
using syringe air.
Sterilizing Handpiece Motor:
After each patient, flush
air by running handpiece motor for 20 seconds.
Remove attachment. Clean all internal surfaces at rear
of handpiece motor with a cotton swab soaked in
isopropyl alcohol until clean. Clean all external surfaces
using a small toothbrush with isopropyl alcohol. Dry
thoroughly. Place handpiece motor and swivel in
sterilization bag – . Follow the sterilization guidelines as
stated by the manufacturer of your sterilizer.
Cleaning-Automated:
Washer-disinfectors are not recommended.
Wrapped Instrument
Sterilization:
Gravity Type Autoclave: Pre-Vacuum Autoclave:
• Minimum of 15 minutes at 132° C • Minimum of 4 minutes at 132° C
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
• dry heat sterilizer
•immersion techniques
• cold sterilizers • ultrasonic cleaners
• sterilizer bags containing disinfectants • hand soaps
• detergents
• chemical disinfectors
•surface sprays
• surgical milk
• surface wipes
Additional Information:
Do not exceed the sterilizer manufacturer’s guidelines for load capacity. Use only paper or paper/plastic bags.
Storage:
Store in bag after sterilization until ready to use.
Dry the motor for 30 minutes minimum using the sterilization dry cycle and allow the motor to cool to room
temperature in bag prior to use. If moisture is present, increase your sterilizer dry cycle appropriately.
Drying:
Containment and transportation:
It is recommended that handpieces are reprocessed as soon as is reasonably practical following use.
Preparation for cleaning:
Remove motor from swivel if applicable or tubing.
Use steam sterilization exclusively for complete sterility as outlined below.
Disinfection:
See Installation, Operating Recommendations, and/or Troubleshooting sections of this manual.
Inspection & function testing:
Single: A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain the
item without stressing the seals.
Packaging:
Manufacturer Contact:
See first or last panels of this manual.
It is the responsibility of the user to use only FDA compliant sterilizing equipment and accessories that are capable of
meeting the manufacturers recommendations below.
CAUTION
NOTICE
Any serious incident occurring from the use of this device should be reported to StarDental immediately.
If incident occurred in a EU Member State the competent authority must be notified.