The
Blis-sonic
SW
requires a separate installation of either a 2/3-line
swivel,
StarDental
Part #261550, or a 4-line swivel,
StarDental
Part #261547,
depending on the handpiece tubing configuration.
1. Grasp the swivel in front of the bearing assembly between the thumb and
index finger as shown.
2. Slide tubing coupler back until the end of the ISO connector is exposed. Hold
in hand as shown with fingers securing the tubing.
3. Align and insert swivel tubes into the ISO connector holes.
4. While applying slight pressure to hold fittings in the connector holes, slide coupler forward and engage threads. Continue threading
clockwise until hand tight against the swivel.
Note:
The swivel tubing connection joint may loosen over time, resulting in water leaks and/or loss of power. Check for tightness
periodically.
Note:
To disconnect scaler from swivel, grasp
scaler firmly with one hand and pull tubing
coupler with the other.
Do not pull on tubing
.
Swivel remains with tubing as shown.
Scaler
Swivel
Tubing Coupler
Tubing
5. Using the universal wrench,
StarDental
Part #203048, hold the
swivel stationary while tightening the tubing coupler until snug.
Be sure to turn tubing coupler, not swivel.
Do not overtighten.
6. Snap scaler onto swivel.
Swivel
Bearing
Assembly
Tubing
ISO Connector
Tubing
Coupler
Wrench
Tubing
Coupler
Turn tubing coupler
while holding
swivel in place
Blis-sonic K
2/3-Line:
StarDental®
Part #264564
Blis-sonic K
4-Line:
StarDental®
Part #264565
Air Pressure: 40 psi (2.7 atmospheres)
at the scaler
Air Flow: 1.5 SCFM (42.0 SLM)
Water Flow: 25 cc/min. @ 15 psi
Length (with tip): 6.7 inches (17.0 cm)
Weight: 1.8 ounces (50.7 grams)
Specifications (Nominal)
Blis-sonic SW
:
StarDental®
Part #264563
(Swivel Sold Separately)
Air Pressure: 40 psi (2.7 atmospheres)
at the scaler
Air Flow: 1.5 SCFM (42.0 SLM)
Water Flow: 25 cc/min. @ 15 psi
Length (with tip): 6.8 inches (17.3 cm)
Weight: 1.9 ounces (54.6 grams)
Sterilizable up to 135°C
135°C
Contents
• Specifications
•
Blis-sonic
SW
&
Blis-sonic
K
Installation
• Tip Installation
• Air Pressure and Water Flow Adjustments
• Tip Operation
• Cleaning &
Sterilization
• Lubricating
• Troubleshooting
• Grip Replacement
• O-Rings & Rear Seal Replacement
• Rotor & Seal Rings Replacement
•
Blis-sonic
K
Water Tube Gasket Replacement
• Replacement Part Numbers
• Warranty
LUBRICATING Blis-sonic
SW and Blis-sonic
K
Always lubricate scaler prior to each autoclave sterilization with DentaLube
®
for Sonic Scalers,
StarDental
Part #261818.
Lubricating
1. To remove
Blis-sonic
K
Scaler from tubing
:
Unthread
from the tubing coupler. Disconnect the
Blis-sonic
SW
from the swivel by grasping the scaler with one hand, pull coupler with other hand.
Do not pull on the tubing
. When separated, the swivel remains with the tubing coupler.
2. Find drive air on
Blis-sonic
K
or notched hole inside of
Blis-sonic
SW
(see illustration).
3. Insert lubricator tip firmly into nozzle. Spray a short burst of
DentaLube
for Sonic Scalers
into
the
Blis-sonic
K
drive air tube or in a notched hole of the
Blis-sonic
SW
.
4. Thread tubing coupler onto the
Blis-sonic
K
or snap
Blis-sonic
SW
onto swivel; attach tip.
Note:
Never run scaler without the tip in place.
5. Run scaler for 20 seconds to circulate lubricant; remove the tip.
6. Steam sterilize, see Sterilization.
Aerosol Lubricant
and Conditioner
DentaLube
for Titan Sonic Scalers
Contains a petroleum lubricant oil (CAS 64742-54-70),
Heptane (CAS 142-82-5), and CO
2
(CAS 124-38-9).
READ BACK PANEL CAREFULLY
.
CFC FREE
To reorder use:
Star
Dental
®
Part #261818
StarDental®
Instrument Solutions
™
DentaLube
for
Titan
®
Sonic Scalers
Blis-sonic
K
Drive Air
Blis-sonic
SW
Notched
Hole
Drive Air
Notched Hole
Use of lubricant other than StarDental “DentaLube for Titan Sonic Scaler” can result in
failure of the scaler.
BEFORE OPERATING THE HANDPIECE, CAREFULLY READ AND FOLLOW ALL OF THESE INSTRUCTIONS. SAVE THESE
INSTRUCTIONS FOR FUTURE REFERENCE AND OBSERVE ALL WARNINGS AND CAUTIONS.
Cleaning and Sterilization
86
6.
DT
E.
IN
FO
(3
83
.4
63
6)
w
w
w
.D
en
ta
lE
Z.
co
m
03
44
06
45
90
J
D
at
e
of
Is
su
e
2
01
9-
10
-0
1
St
ar
D
en
ta
l®
18
16
C
ol
on
ia
l V
ill
ag
e
La
ne
La
nc
as
te
r,
PA
17
60
1
U
.S
.A
.
Pr
in
te
d
in
th
e
U
.S
.A
.
©
20
19
D
en
ta
lE
Z,
In
c.
Em
er
go
E
ur
op
e
Pr
in
se
ss
eg
ra
ch
t 2
0
25
14
A
P
Th
e
H
ag
ue
Th
e
N
et
he
rl
an
ds
EC
RE
P
StarDental®
Instruments
Ca
ut
ion:
F
eder
al la
w r
es
tr
ic
ts
th
is
d
ev
ic
e
to
sale
by
or on the or
der of a den
tist.
Titan®
Blis-sonic
™
Scaler
Instruction Manual
N
on
-S
te
ri
le
Medical Device
Remember:
1) Use the last 1/16” of the tip.
2) Use the edges of the tip,
not
the flats.
3) Keep the tip moving in either of two
motions as shown.
Tip Operation
Edges of
Tip are
Effective
1/16” (1.5 mm)
Effective
Edge
Ineffective
Portion
Flats are
Not
Effective
Effective Scaling Motions
Ineffective Motions
Operation of the
Blis-sonic SW
and K
Ergonomic Sonic Scaler Tip is similar to that of an ultrasonic scaler. Follow these steps to
minimize the risk of damage to restorations.
Universal
-
StarDental
Part #261668
• Use for all surfaces, medium and heavy calculus
Use tips for subgingival scaling and fine access. Select a tip according to the following applications.
ALWAYS USE WATER.
Note:
Normal use results in tip wear.
Replace tips annually.
Use the tip lengths indicated to determine more frequent replacement.
Replace if shorter by .02” (1/2 mm) or more.
1. Always scale with tip at an angle to the restoration margin.
2. Use a “feather touch,” one to two ounces applied pressure.
3. Move tip in a crosshatching, circular or wave fashion (see illustration).
Sickle
-
StarDental
Part #261667
• Recommended for sensitive patients
• Less energetic than the Universal Tip
• Less suitable for deep root scaling
.970 in.
24.6 mm.
.950 in.
24 mm.
Perio
-
StarDental
Part #261669
Perio Left
-
StarDental
Part #263397
Perio Right
-
StarDental
Part #263398
• Recommended for lingual supragingival scaling of
mandibular incisors and deep local pockets
• Provides the same energy as the Universal Tip
• Use to break up heavy calculus deposits
.930 in. (23.6 mm.)
PERIO LEFT
PERIO RIGHT
PERIO
Note:
Use softened and deionized water. Be sure filters operate efficiently. Before installing scaler, thoroughly purge both the dental
unit air line and water line. Verify recommended filtration of 10 microns air and 25 microns water.
If delivery unit has oil/mist lubrication, detach and drain.
Connect the
Blis-sonic
K
by threading to the tubing
coupler.
Push the
Blis-sonic
SW
scaler onto swivel until it
snaps into place.
Blis-sonic SW & K Installation
Blis-sonic
SW
Scaler
Swivel
Tubing Coupler
Tubing Coupler
Blis-sonic
K
Scaler (4-line)
4-line Tubing connector
NOTICE
Do not exceed 40 psi. Patient sensitivity and tooth damage may result.
2. Mount 4/6-line pressure gauge,
StarDental
Part #262591, between the scaler and the air supply tubing for the
Blis-sonic
K Ergonomic Scaler.
Mount the pressure gauge between the swivel and air supply tubing for the
Blis-sonic
SW Ergonomic Scaler.
Snap scaler onto swivel.
3. Turn water OFF.
4. Fully depress foot control. Adjust air supply on
delivery unit until gauge shows 40 psi, 2.72 kg/cm
2
.
1. Thread the selected tip. With the scaler operating, rotate tip a 1/4 turn with fingers.
Air Pressure and Water Flow Adjustments
To prevent damage to the scaler,
do not tighten tips with the universal wrench
. Use the wrench
only to remove tips
as shown in Tip Installation.
5. After adjustment, uncouple the pressure gauge and install scaler directly to the tubing.
6. With scaler running and tip held vertically, adjust water flow until water atomizes at tip. When water
atomizes, the water flow needed is optimized (see illustration above).
Pressure Gauge
NOTICE
NOTICE
H
ygiene
Str
on
g
Do not sharpen or alter tips in any way.
Tip Installation
1. Thread the selected tip onto the scaler with the scaler operating until it stops turning.
Rotate slightly with fingers until tight.
EC
C
S
T
A
Cover
DO NOT USE FOR ORAL SURGERY
2. When the procedure is complete and the scaler is no longer in use, remove the tip and
blow compressed air through the tip to remove water trapped inside the tip. Cap it with
the cover provided,
StarDental
Part #261358. This practice prevents injury to patient or
staff.
3. To change tips, always loosen or tighten with the scaler running. The vibration makes
changing tips easier.
To prevent damage to the scaler,
do not tighten tips with the universal wrench
. Use the universal wrench
ONLY to remove tips
. Use the wrench as illustrated.
StarDental®
Instrument Solutions
StarDental®
Instrument Solutions
Water Tube
Insert Wire
Front
Blis-sonic
K
Blis-sonic
SW
CAUTION
CAUTION
CAUTION
Tip amplitude can exceed 200 µm (.0078 inches).
CAUTION
Do not exceed 135°C
Do not use:
• dry heat sterilizer • cold sterilizers
• ultrasonic cleaners • chemical disinfectors • immersion techniques
• surgical milk • surface wipes • hand soaps • detergents
• surface sprays
• sterilizer bags containing disinfectants
Limits On Reprocessing:
Repeated processing has minimal effect on this
scaler. End of life is normally determined by wear and damage due to use.
Instructions
Scaler, tip, grip, and wrench must be cleaned and sterilized before first use and after each patient.
WARNING
Containment and
transportation:
It is recommended that scalers are reprocessed as soon as is reasonably practical following use.
Preparation for cleaning:
Wear puncture resistant gloves. After each patient, flush air and water by running scaler. Remove tip from
scaler. Lube scaler as described in lubrication.
1. Attach
Blis-sonic
K
to tubing by turning coupler clockwise. Attach
Blis-sonic
SW
by snapping onto swivel.
2. Attach tip to scaler. Never run scaler without tip in place.
3. Run scaler 20 seconds to expel excess lube; remove tip, clean tip per instructions in Cleaning.
4. Detach scaler by reversing
Step 1
.
Cleaning - Automated:
Washer-disinfectors are not recommended.
Cleaning - Manual:
Clean all external surfaces using a small toothbrush with isopropyl alcohol.
Cleaning The Tips
WEEKLY - Blow compressed air through the tip to remove water
trapped inside the tip. Next, clean outside of tip with toothbrush
and isopropyl alcohol. Then run cleaning wire,
StarDental
Part
#256719, through entire tip starting at opening in front
.
Cleaning Blis-sonic SW or K Water Tube
On a regular basis or if a water tube blockage occurs,
remove the tip (and on the
Blis-sonic
SW
also disconnect
from swivel)
.
Push the cleaning wire through the internal
water tube using needle nose pliers at the front on both
the
Blis-sonic
SW
and K
as shown. Push small lengths of
3 to 4 mm at a time until the wire emerges at the opposite
end. Pull wire completely out.
Cleaning the Swivel
Note:
To ensure peak performance and proper functioning of scaler,
clean swivel and mating components daily or as needed. Do not attempt
to disassemble swivel.
1. Disconnect swivel from scaler and tubing. Depress foot control;
verify that water exits tubing.
2. Connect swivel to tubing. Depress foot control; if almost no water
exits front of swivel, clean as follows.
3. Detach swivel from tubing and find water tube.
4. Insert cleaning wire,
StarDental
Part #256719, into water tube.
When the wire will not insert, move wire back and forth to loosen
debris at blockage.
5. Remove wire and apply a blast of air into
water tube with syringe air.
Cleaning Mating Parts of the Swivel
To optimize operation of the swivel, disconnect
mechanism and swab all internal surfaces of
scaler with a cotton swab soaked in isopropyl alcohol until clean. Dry using syringe air.
Clean area in the bearing/spring assembly at rear of swivel using isopropyl alcohol and a toothbrush. Dry using syringe air.
Cleaning Wire
Opening
Bearing/Spring Assembly
Internal Surfaces
Clean
Disinfection:
Use steam sterilization exclusively for complete sterility as outlined below.
Drying:
Dry the scaler for 30 minutes minimum using the sterilization dry cycle and allow the scaler to cool to room
temperature in bag prior to use. If moisture is present, increase your sterilizer dry cycle appropriately.
Lubrication & Maintenance:
See Lubrication section of this manual. The scaler should be routinely maintained, checked for corrosion, worn
parts, etc. and should be replaced on an as needed basis.
Packaging:
Single: A standard sterilization packaging material must be used. Ensure that the pack is large enough to contain
the scaler without stressing the seals.
In sets: scalers may be loaded into dedicated instrument trays or general purpose sterilization trays. Ensure
that cutting edges are protected and do not exceed the tray manufacturer’s limit per tray. Wrap the trays using
appropriate packaging materials and methods.
Wrapped Instrument
Sterilization:
Gravity Type Autoclave: Pre-Vacuum Autoclave:
• Minimum of 15 minutes at 132° C • Minimum of 4 minutes at 132° C
• Minimum of 30 minutes at 121° C
Do not exceed 135° C
Storage:
Store in bag after sterilization until ready to use.
Additional information:
Do not exceed the sterilizer manufacturer’s guidelines for load capacity.
Manufacturer contact:
See front or back panels of this manual.
Inspection & function testing:
See Tip Operation or Troubleshooting sections of this manual.
CAUTION
NOTICE
It is the responsibility of the user to use only FDA compliant sterilizing equipment and accessories that are capable of
meeting the manufacturers recommendations below.
CAUTION
Indications For Use
The scaler is an air driven handpiece which runs at sonic frequency. It effectively removes calculus deposits and stains.
The scaler tips are designed for subgingival scaling and fine access. The Universal tip is for all surfaces and the removal of both heavy and
medium calculus. The Sickle tip is for use on sensitive patients and provides alternate access for difficult posterior areas. ThePerio tip is
for lingual supragingival scaling of mandibular incisors and deep local pockets.
NOTICE
Any serious incident occurring from the use of this device should be reported to StarDental immediately. If incident occurred
in a EU Member State the Competent Authority must be notified.