S.Box Duo S and S.Box Duo ST
Technical characteristics
35
Values determined under ATPD (Ambient Temperature and Pressure, Hygrometry) conditions.
Conditions of use
Pressure range:
700 hPa to 1060 hPa
Temperature:
+5°C to +40°C with side cover
+5°C to +35°C with humidifier
Relative humidity:
Between 15 % and 90 % without condensation
Altitude range:
Approximately 0
2 500 m
Maximum temperature of applied parts:
51°C
Time of contact of the patient with applied
parts:
< 1 minute
Transporting and storage conditions
Pressure range:
700 hPa to 1060 hPa
Temperature:
-25°C to +70°C
Relative humidity:
Up to 90 % without condensation
Electrical characteristics
Device
Input voltage:
24.0 V
20 %
Maximum power consumption:
75 W during a mask disconnection
Current consumption at 20 cmH
2
O with a
4-mm leak:
0.42 A (minimum configuration: S.Box Duo S or
S.Box Duo ST only)
1.99 A (maximum configuration: S.Box Duo S or
S.Box Duo ST with humidifier set to 10 and heated
tube set to 05)
RF emission type (S.Box Duo S or S.Box Duo
ST):
Bluetooth version 2.1+ EDR
Frequency band:
2400 to 2483.5 MHz (ISM band)
Maximum power:
4 dBm
RF emission type (PolyLink):
Bluetooth Smart (BLE 4.1)
Frequency band:
2402 to 2480 MHz (ISM band)
Maximum power:
5.3 dBm
RF emission type (optional S.Box modem):
HSDPA, WCDMA, EDGE, UMTS
Frequency band:
850MHz, 1900 MHz (for US version)
900MHz, 2100MHz (for European version)
800MHz, 850MHz, 2100MHz (for Japanese
version)
Maximum power:
+33 dBm
RF emission type (optional S.Box Wi-Fi
module):
Wi-Fi 802.11b/g/n
Frequency band:
2412 to 2484 MHz (ISM band)
Maximum power:
+18 dBm