7.7
Out Of Range
If measurements are outside the calibrated range, they may seem inaccurate. The out
of range measurements are shown in the OOR file (View Measurement Data Out
Of Range Data). The min. and max. calibration points are given in the RID file, and in
Reconvert Definition Data
See chapter 9 for more.
The user can also define the ranges. Choose the Settings menu, Out Of Range - UV
and DV definition command. A dialog box appears:
Figure 7.7 Set Temperature Out Of Range (OOR) Parameters
Decimal Values
Decimal Values (DV) are raw measurements, sometimes referred to as binary data.
The decimal value is between 0 and 4095, which is the range for the 12 bit AD
define the decimal value range, unless he has obtained these values for that particular
recorder from Star-Oddi.
Unit Values
The decimal values are converted to Unit Values (UV), for example a decimal value of
2000 could give 20°C. The user can select a specified range for temperature and
pressure. If unit values in the data are outside the measurement calibration range of
the recorder, the user has the option to set a max and min range for these unit values.
After selecting the unit value ranges and pressing the OK button, the following
window appears:
Distributed by MicroDAQ.com, Ltd. www.MicroDAQ.com (603) 746-5524