Error message
Error Message/
Error Code
Cause
Coping method
OVER
ERROR
- The weight of the sample to be weighed is in
excess of the maximum capacity.
- Split the sample into several pieces and
weigh them.
- Replace the tare with a lighter one.
- The addition result or calculation result has
exceeded the maximum display digit.
- Clear the calculation result, and then re-
execute the addition/computation while
being careful of the display digit.
UNDER
ERROR
The negative load is below the lower limit.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
Use the dedicated weighing pan and
pan base only.
UC
The weighing accuracy is not guaranteed
because of span calibration setting.
-
-
Select <SELECT MIN 1> at span
calibration.
Select the external calibration weight
which is 1/2 times heavier than the
weighing sample.
DATA MAX
ERROR
Number of the data is over the memory
Clear the data
DISPLAY
ERROR
/ DSP OVER
The calculation result has exceeded the
maximum display digit.
Clear the calculation result, and then
reexecute the computation while being
careful of the display digit.
LOWER
ERROR
The unit/reference weight in Counting/Percentage
mode is below the lower limit.
Choose the samples of which unit
weight/reference weight is larger than
the lower limit.
ERR001
~
ERR099
System error
Record the error code and notify the
store where you purchased the product.
ERR703
- The operation key was pushed at the time of
starting from the standby status.
If the error message is displayed nevertheless the
operation key wasn’t pushed, there is something
wrong with the hardware.
Do not push the operation key while the
balance is in the process of starting
from the standby status.
ERR705
Initial zero adjustment error.
The initial zero adjustment was not completed in
the process of starting from the standby status
because of the unstable load.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
- Check for any wind or vibration.
ERR706
The load is out of the initial zero adjustment
range.
- Do not put any load on the weighing pan
at the power-on of the balance.
ERR709
ERR710
ERR711
-
-
The load is unstable at the zero adjustment/tare
subtraction.
Span calibration time-out error.
-
-
Improper setting of the weighing pan or
pan base is suspected.
Check for contact with other object.
- Check for any wind or vibration.
ERR717
The mass of the calibration weight is 1% differ
from the designated mass at the external span
calibration.
Check the calibration value of the
weight and use the proper calibration
weight.
ERR718
The mass of the calibration weight is under 50%
of the maximum capacity at “span calibration” by
external calibration weight.
Use the calibration weight of which
weight is equal to the maximum
capacity.
ERR719
The adjust value by “external span calibration” is
over 1% of the maximum capacity.
- Check the mass of the weight used for
the external span calibration.
ERR722
Tare key is pushed during the Preset tare
operation.
Do not push the Tare key during the
Preset tare operation.
Содержание MG-S Series
Страница 2: ......
Страница 10: ......
Страница 67: ...MG S series operation manual 6 External input output functions 57 ...
Страница 68: ...6 External input output functions MG S series operation manual 58 ...
Страница 84: ...8 Controlling and adjustment functions MG S series operation manual 74 ...
Страница 103: ...MG S series operation manual Appendix 93 Appendix 2 Dimensional outline drawing MG S322 MG S1501 MG S8200 ...
Страница 109: ...MG S series operation manual Appendix 99 Appendix 7 Windshield assembly instructions ...
Страница 110: ...Appendix MG S series operation manual 100 ...
Страница 111: ...MG S series operation manual Appendix 101 ...
Страница 113: ......
Страница 114: ......