This guide will assist you with manual programming of the Bed-Check

®

 Model Vr control unit. More 

detailed manual programming instructions can be found in the Model Vr User Guide, and more 

extensive programming can be accomplished with the use of a PC.

For programming, only the control unit and power cord are needed. Neither the Sensormat

®

 pad nor the nurse 

call cord are needed during programming. At each step, if no change is needed, simply press RESET to skip that 

step and advance to the next one.

ARMING/DISARMING THE SYSTEM

Press and release the RESET button to go from MONITOR to HOLD, or from HOLD to MONITOR, or to silence an 

alarm. Please note that if the system is on HOLD, it will reset itself and resume monitoring if weight is placed 

back on the Sensormat pad. In addition, if the RESET button is pressed to put on HOLD and weight remains on 

the Sensormat pad, the unit will resume monitoring in 25 seconds.

Before you begin PROGRAMMING, the unit must be put in the Programming Mode.

 

Plug the unit in and set the 

delay time display on the front panel to 9 using the 

 

    

button. Unplug the power cord (either end) and wait 2-3 

seconds. Press and hold the RESET button on the control unit and reconnect the power cord. Continue to hold 

the RESET button until you hear a beep and all lights in the status window light up as shown in Step 1 below.

1.  AlARM SOuND

 

(bypass this step by pressing reset)

Use the

 

     

button to change the display number to coordinate with the sound you 

would like.
 

0.  Fast Beep 

4.  Warble

 

1.  Medium Beep 

5.  Triple Tone

 

2.  Slow Beep 

6.  Music

 

3.  Ring

After selection, press the RESET button to continue to Step 2.

2.  NuRSE cAll SYSTEM RElAY

 

(bypass this step by pressing reset)

This sets up the interface compatibility between the control unit and your nurse call 

system. It is rarely changed, with the most common settings being either 0 or 3. If 

unsure, check with your BioMed Department.
 

0.  Continuous Closure (most common)

 

1.  Repeated Pulse

 

2.  One Shot

 

3.  No Pull-Out protection (for no nurse call connection)

Press RESET to move to Step 3.

Before setting

You must see

ALARM

HOLD

MON

Before setting

You must see

HOLD

MON

Quick Programming Guide

Model Vr