This guide will assist you with manual programming of the Bed-Check
®
Model Vr control unit. More
detailed manual programming instructions can be found in the Model Vr User Guide, and more
extensive programming can be accomplished with the use of a PC.
For programming, only the control unit and power cord are needed. Neither the Sensormat
®
pad nor the nurse
call cord are needed during programming. At each step, if no change is needed, simply press RESET to skip that
step and advance to the next one.
ARMING/DISARMING THE SYSTEM
Press and release the RESET button to go from MONITOR to HOLD, or from HOLD to MONITOR, or to silence an
alarm. Please note that if the system is on HOLD, it will reset itself and resume monitoring if weight is placed
back on the Sensormat pad. In addition, if the RESET button is pressed to put on HOLD and weight remains on
the Sensormat pad, the unit will resume monitoring in 25 seconds.
Before you begin PROGRAMMING, the unit must be put in the Programming Mode.
Plug the unit in and set the
delay time display on the front panel to 9 using the
button. Unplug the power cord (either end) and wait 2-3
seconds. Press and hold the RESET button on the control unit and reconnect the power cord. Continue to hold
the RESET button until you hear a beep and all lights in the status window light up as shown in Step 1 below.
1. AlARM SOuND
(bypass this step by pressing reset)
Use the
button to change the display number to coordinate with the sound you
would like.
0. Fast Beep
4. Warble
1. Medium Beep
5. Triple Tone
2. Slow Beep
6. Music
3. Ring
After selection, press the RESET button to continue to Step 2.
2. NuRSE cAll SYSTEM RElAY
(bypass this step by pressing reset)
This sets up the interface compatibility between the control unit and your nurse call
system. It is rarely changed, with the most common settings being either 0 or 3. If
unsure, check with your BioMed Department.
0. Continuous Closure (most common)
1. Repeated Pulse
2. One Shot
3. No Pull-Out protection (for no nurse call connection)
Press RESET to move to Step 3.