Merlin™ PSA Help Manual
3-5
Interval and Refractory Markers (Full Markers)
Interval and refractory markers are shown in the following diagram.
Figure 3-1. Interval and refractory markers (full markers)
Special Event Markers
Marker
Description
Example
--> A-Noise
or --> V-
Noise
Entry into noise
reversion mode
A-noise or V-
Noise
Continuation of
noise reversion
mode
<-- A-Noise
or <-- V-
Noise
Exit from noise
reversion mode
Burst---
Burst stimulation
Table 3-2. Special event markers
1
6
4
914
907
A-V Interval
A-A Interval
V-V Interval
Refractory
Period (Line)
Absolute Refractory
Period (Thicker Line)