Alarms
The Alarms page allows you to set lower and upper limits and assign actions to readings for the
sensor.
For the desired gas type, enter a Lower Limit, an Upper Limit, select an Action from the
dropdown menu, and click “Submit” to save that information.
When connected to a digital card, if desired, one of the two relays (or both simultaneously) can
be energized. There are four possible actions for the alarms:
“None” will not energize any relays.
“AL1” will energize Relay 3;
“AL2” will energize Relay 4;
“Both” will energize Relays 3 and 4.