10.3.2
Clean the inside of the valve body
1.
After taking out the rotor, wet a cotton stab with methanol or isopropyl alcohol.
2.
Clean gently by dabbing the inside of the valve body to remove any residue.
3.
Also, be careful to remove any fiber left during the cleaning.
4.
Visually check the inside of the valve body. The conical surface must be smooth. The presence of a
scratch between the holes can lead to leakage, in this case the valve body must be replaced.
10.3.3
Clean the rotor
1.
Carefully grasp the rotor tip, taking care not to damage the metal or polymer.
2.
Immerse it briefly in the solvent (methanol or isopropyl alcohol).
3.
Carefully wipe the polymer with a clean tissue and make sure that no fiber is left by the tissue.
4.
Inspect the rotor. If it has scratches or shrinkage, it must be replaced.
Dirty rotor after several
months of analyses.
Original rotor