6.6
Options
The
Method\Option
menu allows you to automate several injection cycles. Note that between the
injection cycle (analysis) and the next storage cycle, there is no washing cycle.
The 'Cycle number' parameter allows you to define the number of storage and analysis cycles you wish to
perform.
Usually in the case of multi-cycle, not all loops are used. It is important to set up the TGA analysis cycle
correctly so that the loops used by the storage sequence are injected and analyzed. It is also important to
check the box 'Running by external start' so that the IST16 waits for the signal from the TGA at the end of
the analysis of the last loop to restart the next storage cycle.
The checkbox 'Run the inject sequence after the storage sequence' allows to start the injection cycle as
soon as the storage cycle is finished.
If the 'Running by external start' checkbox is enabled and you start the storage cycle by pressing the 'Start'
button, the IST16 waits for the TGA to send a Start request to start the storage cycle.
7.
Gas sampling valve and Multi-injection modes
Both modes can be accessed via the
Interface\Configuration\Advanced options
menu.
With these two modes, the interface allows you to store and then inject a single loop several times in a row
for a set period of time. As soon as the loop is injected, it returns to the storage position.
There are two ways of working, which are described in the following paragraphs.