BASIC SYSTEM OPERATION 

 
SETTING SYSTEM ZERO 

Make sure the scale is free and clear of any obstructions and press and hold the 

ZERO / 

WEIGH

  button.    The  displayed  message  will  indicate 

HOLD  TO  ZERO

”  and  count 

down to zero.  

Make sure that nothing is in contact with the scale while zeroing the system.  

In a few seconds, the display will read “

WT = 0.0 Lb

  (or

  Kg

)

.  This action also resets   

previously stored weight, height, and BMI values to zero.   

 

 
 
 

 

 

 

WEIGHING

 

Position  the  patient  on  the  scale.    The  care-giver  must  be  standing  on  the  scale  when 

moving  the  wheelchair  on  or  off  the  scale.  Lock  the  patient’s  wheelchair  to  prevent 

movement.  Press  the 

ZERO/WEIGH

  button.  The weight  stable indicator

 

” flashes 

on the display.  When the weight is stable, the weight stable indicator remains solid.  The 
display will indicate the patient’s weight in either pounds or kilograms; example: “

WT = 

123.5 Lb

”.  The stable weight is auto stored in memory. 

NOTE:  

If patient will be using a cane for support on the scale, place cane on scale while 

zeroing the system.  This will ensure that the patient’s NET weight will be displayed.  It is 
recommended that the system be zeroed prior to each new patient.  

 

 

CALCULATING BMI

 

Press to calculate BMI.   
When the 

BMI

 button is pressed, the default starting value is set either to 65 in or to 165 

cm; example:  “

HT = 65 in

”.  Use the 

UP

 or 

DOWN

 arrows to adjust the default height to 

the value of the patient’s height.  

 

 

NOTE: 

  To  calculate  the  BMI,  a  patient’s  stable  weight  needs  be  stored  as  indicated  in

 

WEIGHING

” above, if no stable weight has not been stored “

NO  WEIGHT  DATA

” 

will be displayed and the display will then go back to the weigh screen “

WT = 0.0 Lb

”. 

 

 

Press 

ENTER

  to  save  height  data  and  display  BMI  calculation.    The  BMI  will  be 

displayed; example: “

BMI = 20.5

”. 

CARE-GIVER MUST ASSIST WHEELCHAIR-BOUND PATIENTS.  

DO NOT LEAVE PATIENT UNATTENDED ON THE SCALE PLATFORM.  

 

WARNING 

Continued next page 

Содержание SR735i

Страница 1: ..._MICROBAN_201210 Page 1 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S...

Страница 2: ...USE 6 MICROBAN ANTIMICROBIAL PROTECTION 6 STORAGE 6 CLEANING AND DISINFECTING 7 SPECIFICATIONS 8 BUTTON FUNCTIONS 9 BASIC SYSTEM OPERATION 11 BATTERY REPLACEMENT 12 THEORY OF OPERATION 13 CALIBRATION...

Страница 3: ...wanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc PACKING CHECKLIST SR735i Platform Scale DESCRIPTION QUANTITY Box 1 WHE...

Страница 4: ...lease call the service hotline at 1 800 654 6360 STEP 2 Verify that the serial number on the Display Unit matches that on the Wheelchair Scale Assembly STEP 3 The SR735i Display may be wall mounted Fi...

Страница 5: ...ing Bracket Figure 4 Use the wall mounting hardware provided for installing the bracket STEP 2 Carefully attach the printer to the Printer Mounting Bracket Connect the printer cable from the printer t...

Страница 6: ...be viewed either in pounds or kilograms with a displayed resolution of 0 1 for each INTENDED USE The SR735i Wheelchair Scale System is specifically designed for weighing both motorized wheelchair and...

Страница 7: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Страница 8: ...e AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds RECALL Recalls last stored stable weight height and BMI AVERAGING Automatic digital filter POWER SUP...

Страница 9: ...when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and BMI Also indicated on display when...

Страница 10: ...ON TIME Use UP or DOWN arrow buttons to adjust the ON TIME The ON TIME may be set from 30 to 300 seconds in 30 second increments Press ENTER to save changes Setting TIME and DATE Use the UP arrow butt...

Страница 11: ...When the weight is stable the weight stable indicator remains solid The display will indicate the patient s weight in either pounds or kilograms example WT 123 5 Lb The stable weight is auto stored i...

Страница 12: ...example RCL WT 123 5 Lb RCL HT 65 in and RCL BMI 20 5 BATTERY REPLACEMENT STEP 1 The display will read REPLACE BATTERY STEP 2 Figure 7 Unscrew the panel screw on the Battery Compartment Cover and remo...

Страница 13: ...plifier is configured to current sum the output of each cell with potentiometers serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer pro...

Страница 14: ...subset of weight that is less than half of the combined weight DO NOT USE barbells or uncalibrated weights STEP 2 Zero the scale by pressing and holding ZERO WEIGH button STEP 3 Place the subset of c...

Страница 15: ...weight of the calibrated weights Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 500 00 Lb is...

Страница 16: ...EP 5 Switch the scale system out of the Calibration Mode on the display board Figure 9 STEP 6 Place the left end cover on the display housing Attach using three 3 cover screws EHR OUTPUT Electronic Me...

Страница 17: ...ry Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label is c...

Страница 18: ...abilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no...

Страница 19: ...and Service Manual Part No MAN735i _MICROBAN_201210 Page 19 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments co...

Страница 20: ...AN735i _MICROBAN_201210 Page 20 of 20 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Отзывы: