SR Instruments SRScales SRV947R Скачать руководство пользователя страница 5

Model SRV947R-BT 24 X 48 Remote Display Platform 
Operating and Service Manual 
Part No. MAN947R-BT_210610 

 

Page 5 of 18 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2021

 

S

Instruments, Inc. 

CLEANING and DISINFECTING 

 

CLEANING 

To clean the display / user interface and other scale contact areas: 

 

Use a soft cloth dampened with water and mild detergent to clean scale surfaces. 

 

Wipe surface with clean soft cloth dampened with water and then dry with clean soft cloth.  

 

Do  not  use  abrasive  materials  to  clean  scale  surface  to  prevent  damage  to  the  surface 
finish. 

 

Do not spray liquid directly onto scale surfaces.  Use only a damp cloth. 

 

DISINFECTION 

To disinfect the display / user interface and other scale contact areas: 

 

Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth.  Cloth 

cannot  be  dripping  wet.    Follow  manufacturer’s  instruction  on  the  proper  use  of 
commercially available disinfectants.  

 

Disinfectant solutions with 1% sodium hypochlorite or 70% isopropyl alcohol are suitable 
for display / user interface and other scale contact surfaces. 

 

After  disinfecting,  use  a  soft  cloth  dampened  with  clean  water  and  dry  with  a  soft  clean 
cloth to prevent buildup of material on scale finish. 

 

Do  not  use  abrasive  material  to  disinfect  /  clean  scale  surfaces  to  prevent  damage  to  the 
surface finish. 

 

Do not spray liquid directly onto scale surfaces.  Use only a damp cloth.  

 

WARNING:

 

DO  NOT  SPRAY  CLEANING  SOLUTION  OR  LIQUIDS  DIRECTLY 
ON SURFACES TO BE CLEANED

 

WARNING:  EXPOSURE  TO  EXCESSIVE  LIQUID  WILL  DAMAGE  USER 

INTERFACE KEYPAD. 

 

WARNING:

 

DO  NOT 

USE  PRESSURIZED  WATER  OR  STEAM.  THE  SCALE 

SYSTEM CONTAINS ELECTRONIC COMPONENTS THAT MAY BE 

ADVERSELY  AFFECTED  BY  EXPOSURE  TO  SUCH  AN 

ENVIRONMENT. 

Содержание SRScales SRV947R

Страница 1: ...610 Page 1 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc SRV947R BT REMO...

Страница 2: ...F CONTENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV947R BT 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUN...

Страница 3: ...rsonnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are...

Страница 4: ...patient s weight The patient s weight is displayed on an LCD screen With a push of a button weight data may be viewed in either pounds or kilograms INTENDED USE The SRV947R BT Portable Platform Scale...

Страница 5: ...act areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available...

Страница 6: ...matically turn on and lock in on a stable weight AVERAGING Automatic digital filter POWER SUPPLY AC Wall Power Supply Medical Grade CALIBRATION Calibration is traceable to NIST standards OPERATING CON...

Страница 7: ...ts com Copyright 2021 SInstruments Inc MODE ZERO HOLD RECAL L BUTTON FUNCTIONS Figure 1 SRV947R BT Button Label The ZERO WEIGH button is used to clear the Recall memory and start a new weighing cycle...

Страница 8: ...en pounds or kilograms The system will remain in the same mode as last used unless manually changed WEIGH The WEIGH button is used to clear the Recall memory and start a new weighing cycle HOLD RECALL...

Страница 9: ...or remote wireless operation Use your Android device to display weight and control the SRV947R BT scale functions Figure 2 Ensure your remote display indicates CONNECTED before use The WEIGH button wi...

Страница 10: ...lified by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor ov...

Страница 11: ...semiconductors on the printed circuit boards may be damaged by electrostatic discharge ESD Be sure to use proper handling precautions at all times IMPORTANT CALIBRATION CHECK Qualified service person...

Страница 12: ...r a parameter saved NOTE Ensure that nothing is in contact with the scale system during this procedure Remove hands from the system when noting the displayed calibration results CAL MENU data Displays...

Страница 13: ...the digits using the Right Button until desired value is shown Press the Left Button and continue the same with the remaining digits d Press the Left Button to save or Right Button to toggle to quit S...

Страница 14: ...Press the Left Button again to save and the display will show SAV d Calibrate using the Three Point Calibration Curve zero half scale and full scale NOTE The scale can also be calibrated with a Two p...

Страница 15: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clean underneath stand on each of the four corners to see if one corner is not weighing correctly System fails to pe...

Страница 16: ...rm Operating and Service Manual Part No MAN947R BT_210610 Page 16 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrume...

Страница 17: ...ay lay flat on a desk or end table If desired provisions for wall mount have been incorporated and can be accomplished in the manner described below A The display should be installed by qualified pers...

Страница 18: ...ual Part No MAN947R BT_210610 Page 18 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInst...

Отзывы: