SR Instruments SRScales SRV947R Скачать руководство пользователя страница 3

Model SRV947R-BT 24 X 48 Remote Display Platform 
Operating and Service Manual 
Part No. MAN947R-BT_210610 

 

Page 3 of 18 

 
 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2021

 

S

Instruments, Inc. 

PACKING CHECKLIST – Model SRV947R-BT 

Remote Display Platform Scale 

 

 

 

ASSEMBLY 

 

STEP 1

:  Unpack  the  scale  system  and  check  parts  against  the 

PACKING  CHECKLIST

.    If 

there are any missing or damaged parts, call the Service Hotline at:  1-800-654-6360.   

STEP 2

:  Determine installation location for scale and display, keeping in mind the display cable 

length is six (6) feet and the display power cord length is four (4) feet. The display may lay flat on 
a  desk  or  end  table.  If  desired,  provisions  for  wall  mount  have  been  incorporated,  and  can  be 
accomplished in the manner described below. 

A.

 

The  display  should  be  installed  by  qualified  personnel.  The  mounting  hardware  provided 
will cover most wall applications. 

B.

 

Determine a position on the wall comfortable to the user.  Mark the location of the two (2) 
mounting  holes  on  the  wall  (holes  are  spaced  2.5”  on  center  apart)  using  the  template 
provided on page 16.  

C.

 

Drill the two (2) pilot holes, no larger than ¼” diameter in the locations marked. Install the 
two (2) self-drilling screw anchors. 

D.

 

Install the two (2) #8 x 1” flat head screws supplied, leaving approximately ¼” projecting 
from the wall. 

STEP 3

:  Connect signal cable to the underside of the platform. 

STEP 4

:  Place SRV947R-BT scale system on a flat, hard floor surface. Ensure all six feet of the 

scale are in contact with the floor and scale does not rock

.  

NOTE:

 Scale should not be placed on carpet or uneven floor surfaces as this may interfere with 

scale accuracy. 

STEP 5

:  Connect the display cable to the scale, then connect the power cable to the display and 

plug power adapter into wall. 

 

DESCRIPTION 

QUANTITY 

 

SCALE BASE  

1 ea. 

 

SCALE DISPLAY  

 1 ea. 

 

MOUNTING HARDWARE 

1 ea. 

 

AC WALL POWER SUPPLY (MEDICAL GRADE) 

1 ea. 

 

CERTIFICATE OF CALIBRATION 

1 ea. 

 

MANUAL 

1 ea. 

Содержание SRScales SRV947R

Страница 1: ...610 Page 1 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc SRV947R BT REMO...

Страница 2: ...F CONTENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV947R BT 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUN...

Страница 3: ...rsonnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are...

Страница 4: ...patient s weight The patient s weight is displayed on an LCD screen With a push of a button weight data may be viewed in either pounds or kilograms INTENDED USE The SRV947R BT Portable Platform Scale...

Страница 5: ...act areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available...

Страница 6: ...matically turn on and lock in on a stable weight AVERAGING Automatic digital filter POWER SUPPLY AC Wall Power Supply Medical Grade CALIBRATION Calibration is traceable to NIST standards OPERATING CON...

Страница 7: ...ts com Copyright 2021 SInstruments Inc MODE ZERO HOLD RECAL L BUTTON FUNCTIONS Figure 1 SRV947R BT Button Label The ZERO WEIGH button is used to clear the Recall memory and start a new weighing cycle...

Страница 8: ...en pounds or kilograms The system will remain in the same mode as last used unless manually changed WEIGH The WEIGH button is used to clear the Recall memory and start a new weighing cycle HOLD RECALL...

Страница 9: ...or remote wireless operation Use your Android device to display weight and control the SRV947R BT scale functions Figure 2 Ensure your remote display indicates CONNECTED before use The WEIGH button wi...

Страница 10: ...lified by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor ov...

Страница 11: ...semiconductors on the printed circuit boards may be damaged by electrostatic discharge ESD Be sure to use proper handling precautions at all times IMPORTANT CALIBRATION CHECK Qualified service person...

Страница 12: ...r a parameter saved NOTE Ensure that nothing is in contact with the scale system during this procedure Remove hands from the system when noting the displayed calibration results CAL MENU data Displays...

Страница 13: ...the digits using the Right Button until desired value is shown Press the Left Button and continue the same with the remaining digits d Press the Left Button to save or Right Button to toggle to quit S...

Страница 14: ...Press the Left Button again to save and the display will show SAV d Calibrate using the Three Point Calibration Curve zero half scale and full scale NOTE The scale can also be calibrated with a Two p...

Страница 15: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clean underneath stand on each of the four corners to see if one corner is not weighing correctly System fails to pe...

Страница 16: ...rm Operating and Service Manual Part No MAN947R BT_210610 Page 16 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrume...

Страница 17: ...ay lay flat on a desk or end table If desired provisions for wall mount have been incorporated and can be accomplished in the manner described below A The display should be installed by qualified pers...

Страница 18: ...ual Part No MAN947R BT_210610 Page 18 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInst...