THEORY OF OPERATION 

 

SR Instruments patient weighing systems are digital scales.  Strain-gauge force cells convert the force 
of an applied weight into an analog signal.  This signal is amplified by an operational amplifier and 
converted to a digital signal by analog to digital converter.  The digital signal is transferred to a micro-
controller where it is filtered, converted to appropriate units, and displayed on a liquid crystal display. 
Strain-gauge  force  cells  each  contain  four  strain  gauges  mounted  in  a  full  Whetstone-bridge 
configuration.  These bridges convert the physical movement of the force cell, due to the applied mass 
on  the  system,  into  minute  changes  in  electrical  resistance.    These  changes  in  resistance  produce  a 
voltage difference across the Whetstone-bridge, which is amplified by the operational amplifier.   
The output of the operational amplifier is digitized by the analog to digital converter.  The converter 
integrates  the  analog  signal  onto  the  integrating  capacitor  over  a  short  interval.    The  integrating 
capacitor is then discharged  at  a  rate proportional  to the  reference voltage  applied to the  converter.  
The residual voltage on the integrating capacitor is then multiplied by a factor and again discharged at 
a  rate  proportional  to  the  reference  voltage.    The  residual  voltage  from  this  discharge  is  again 
multiplied by a factor and again discharged.  The time taken to discharge the capacitor is proportional 
to the voltage from the operational amplifier, which is proportional to the applied load on the force 
cells.  The time is stored as a binary number in the analog to digital converter and is transferred to the 
micro-controller when the conversion is complete. 
The micro-controller averages and filters the digital output of the analog to digital converter, subtracts 
the value saved during the system zero operation and scales the filtered output, then displays the result 
on the liquid crystal display.  The micro-controller performs a rolling average of data for continuous 
weigh and averages the data before locking in on the reading. 

Содержание SRScales SRV947R

Страница 1: ...610 Page 1 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInstruments Inc SRV947R BT REMO...

Страница 2: ...F CONTENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV947R BT 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUN...

Страница 3: ...rsonnel The mounting hardware provided will cover most wall applications B Determine a position on the wall comfortable to the user Mark the location of the two 2 mounting holes on the wall holes are...

Страница 4: ...patient s weight The patient s weight is displayed on an LCD screen With a push of a button weight data may be viewed in either pounds or kilograms INTENDED USE The SRV947R BT Portable Platform Scale...

Страница 5: ...act areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available...

Страница 6: ...matically turn on and lock in on a stable weight AVERAGING Automatic digital filter POWER SUPPLY AC Wall Power Supply Medical Grade CALIBRATION Calibration is traceable to NIST standards OPERATING CON...

Страница 7: ...ts com Copyright 2021 SInstruments Inc MODE ZERO HOLD RECAL L BUTTON FUNCTIONS Figure 1 SRV947R BT Button Label The ZERO WEIGH button is used to clear the Recall memory and start a new weighing cycle...

Страница 8: ...en pounds or kilograms The system will remain in the same mode as last used unless manually changed WEIGH The WEIGH button is used to clear the Recall memory and start a new weighing cycle HOLD RECALL...

Страница 9: ...or remote wireless operation Use your Android device to display weight and control the SRV947R BT scale functions Figure 2 Ensure your remote display indicates CONNECTED before use The WEIGH button wi...

Страница 10: ...lified by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor ov...

Страница 11: ...semiconductors on the printed circuit boards may be damaged by electrostatic discharge ESD Be sure to use proper handling precautions at all times IMPORTANT CALIBRATION CHECK Qualified service person...

Страница 12: ...r a parameter saved NOTE Ensure that nothing is in contact with the scale system during this procedure Remove hands from the system when noting the displayed calibration results CAL MENU data Displays...

Страница 13: ...the digits using the Right Button until desired value is shown Press the Left Button and continue the same with the remaining digits d Press the Left Button to save or Right Button to toggle to quit S...

Страница 14: ...Press the Left Button again to save and the display will show SAV d Calibrate using the Three Point Calibration Curve zero half scale and full scale NOTE The scale can also be calibrated with a Two p...

Страница 15: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clean underneath stand on each of the four corners to see if one corner is not weighing correctly System fails to pe...

Страница 16: ...rm Operating and Service Manual Part No MAN947R BT_210610 Page 16 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrume...

Страница 17: ...ay lay flat on a desk or end table If desired provisions for wall mount have been incorporated and can be accomplished in the manner described below A The display should be installed by qualified pers...

Страница 18: ...ual Part No MAN947R BT_210610 Page 18 of 18 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2021 SInst...

Отзывы: