SR Instruments SR300 Скачать руководство пользователя страница 1

Model – 3000 Gram x 1 Gram Scale 
Instruction Manual - S/N 1000+ 
Part No.:  MAN315_110121 

 

Page 1 of 2 

S

Instruments, Inc., 

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

 

 

email:  

[email protected]

     

  

Copyright 2011

 

S

Instruments, Inc. 

 

SPECIFICATIONS 

 

Max. capacity: 

3,000g/105oz 

Min. capacity: 

5g/0.2oz 

Division:   1g/0.1oz 
Power source: 

Two (2) CR2032 lithium 
batteries 

Accuracy: 

0.2% +/- 1 digit of 
displayed resolution for 
calibrated range 

 

FEATURES 

 

Portability:   

Small and compact 

Tare: 

  Single 

button 

Two units: 

 

Gram or ounce  

Auto power off: 

120 seconds 

Warranty: 

 

Two (2) years 

 

BATTERY INSERTION/ 
REPLACEMENT 

1.

 

Remove the Battery Compartment Cover by 
pressing the tab towards the cover and lifting it 
off (Figure 2). 

2.

 

Insert two (2) 3-Volt lithium batteries (CR2032) 
with battery “+” in the up position. 

3.

 

Replace the Battery Compartment Cover. 

 

TO OPERATE SCALE 

1.

 

Place scale on a flat surface in a sheltered area. 

2.

 

Press “

ON/OFF

” button to zero (“

0”) 

scale. 

3.

 

Press “

Kg / Lb

” button to select between grams, 

kilograms, ounces, or pounds and ounces.   

4.

 

Scale is ready for use.    

 

USING THE TARE FUNCTION 

1.

 

Place scale on a flat surface in a sheltered area. 
Press the “

ON/OFF

” button to turn the scale on.  

2.

 

When “

0

” appears place the empty container 

onto the weighing platform. 

3.

 

Press “

TARE

” button and display will return to 

zero, eliminating the weight of the container. 

4.

 

Place object to be weighed into the container to 
obtain the net weight. 

5.

 

Press “

TARE

” button again to exit. 

 

 

 

 

MANUAL POWER OFF 

After use, turn the scale off by pressing the 

ON/OFF

” button to prolong battery life. 

 

AUTO POWER OFF FEATURE

 

The scale is also equipped with an “

AUTO 

POWER OFF

” function. It will turn off in 

approximately 120 seconds. 
 

HOLD FUNCTION 

1.

 

To temporarily retain weight. Place item on the 
scale. Press the “

HOLD

” button once.  

2.

 

Remove the item from the scale. The held 
weight will remain on the scale display for 5 
seconds. 

 

CAUTION TO BE EXERCISED 

1.

 

Place the scale in a sheltered area when in use.  
Air movement on the scale may affect accuracy.   

2.

 

Always place object onto the platform gently. 

DO NOT

 weigh objects exceeding the full 

capacity as this may permanently damage scale. 
An “

o-Ld

” message will appear on the display if 

the scale is over-loaded.   

3.

 

Replace with two (2) CR3032 lithium batteries 
when battery symbol 

 appears (Figure 2). 

 

CALIBRATION METHOD

  

1.

 

Calibration switch is located under “

VOID

” 

sticker; use a small screw driver to slide switch, 
inside scale to Cal mode. 

2.  A random number will be displayed. Press the 

ON / OFF

” button to display “

3000gr

”.  

3.  Place a 3000 gram calibrated weight on scale. 
4.   Press “

ON / OFF

” button. The scale will 

display “

CAL

” and then “

PASS

”. 

 

NOTE:

  If “

o-Ld

” appears; recalibrate. 

5.  Remove weight. Slide calibration switch out of 
 

Cal  mode. 

6.

 

Place 3000 gram weight back on scale to verify 
calibration.  

Отзывы: