CLEANING and DISINFECTING
CLEANING
Use water and mild detergent to clean scale surfaces.
Do not use abrasive material to disinfect / clean scale surfaces to prevent damage to the
surface finish.
To clean the display / user interface:
Wipe surface with clean soft cloth dampened with water and then dry with clean soft cloth.
Do not spray liquid directly onto interface surfaces. Use only a damp cloth.
To clean platform / patient contact areas:
Use a soft bristle brush with mild detergent and water when cleaning platform / patient
contact areas. Pat dry with a clean cloth, or allow to air dry.
DISINFECTION
Follow manufacturer’s instruction on the proper use of commercially available
disinfectants.
Disinfectant solutions with 1% sodium hypochlorite or 70% isopropyl alcohol are suitable
for display / user interface and other scale contact surfaces.
To disinfect the display / user interface:
Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth. Cloth
cannot be dripping wet.
After disinfecting, use a soft cloth dampened with clean water and dry with a soft clean
cloth to prevent buildup of material on display.
Do not spray liquid directly onto interface surfaces. Use only a damp cloth.
To disinfect platform / patient contact areas:
Apply disinfectant solution to platform / patient contact areas.
Patient contact area should remain wet with disinfectant solution for six minutes. Pat dry
with a clean cloth, or allow to air dry.
WARNING: EXPOSURE TO EXCESSIVE LIQUID WILL DAMAGE USER
INTERFACE KEYPAD.
WARNING:
DO NOT
USE PRESSURIZED WATER OR STEAM. THE SCALE
SYSTEM CONTAINS ELECTRONIC COMPONENTS THAT MAY BE
ADVERSELY AFFECTED BY EXPOSURE TO SUCH AN
ENVIRONMENT.