Model SR7005
i
Wall Mount In-Room Mini Scale
Operating and Service Manual
Part No. MAN7005i_200520
Page 14 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2020
S
Instruments, Inc.
TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
The characters only appear on half of the
display.
Press the
WEIGH
button or remove one battery. Wait
five seconds, then re-install battery and try the
WEIGH
button again.
The display lights appear to work, but do not
respond to button activation.
Check to ensure the membrane switch label is correctly
plugged into display board.
Check to ensure the calibration switch is not in the
Calibration Mode (Figure 5).
The display shows no reading at all.
Check to ensure batteries are installed correctly (see
directions for
BATTERY REPLACEMENT
). Check
display cable to make sure it is connected securely
For additional information or assistance, telephone our Service Hotline: 1-800-654-6360