THEORY OF OPERATION 

 

SR  Instruments  patient  weighing  systems  are  digital  scales.  Strain-gauge  force  cells  convert  the 
force  of  an  applied  weight  into  an  analog  signal.  This  signal  is  amplified  by  an  operational 
amplifier  and  converted  to  a  digital  signal  by  an  on-chip  analog  to  digital  converter.  The  digital 
signal is filtered, converted to appropriate units, and displayed on the liquid crystal display. 
 
Strain-gauge  force  cells  each  contain  four  strain  gauges  mounted  in  a  full  Wheatstone-bridge 
configuration. These bridges convert the physical movement of the force cell, due to the applied 
mass  on  the  system,  into  minute  changes  in  electrical  resistance.  These  changes  in  resistance 
produce a voltage difference across the Wheatstone-bridge, which is amplified by the operational 
amplifier. The amplifier is configured to current sum the output of each cell, with potentiometers 
serving  to  normalize  the  sensitivity  (voltage  out  per  unit  of  weight  applied)  of  each  bridge.  The 
offset  potentiometer  produces  a  small  current,  which  nulls  the  output  of  the  amplifier  for  an 
unloaded system.  
 
The output of the operational amplifier is digitized by the analog to digital converter. The sigma-
delta  converter  sums  a  rapid  sequence  of  0's  (0  volts)  and  1's  (reference  voltage)  to  achieve 
balance with the input signal from the amplifier.  
 
The micro-controller filters the digital output of the analog to digital converter, subtracts the value 
saved during the system zero operation and scales the filtered output, and then displays the result 
on  the  liquid  crystal  display.  The  micro-controller  performs  a  moving-median  filter  of  data  for 
continuous  weigh  and,  for  AutoHold;  the  micro-controller  performs  checks  for  signal  stability 
before locking in on the reading. If the data variance is greater than 0.1% in the AutoHold mode, 
the micro-controller will reset the filter and start a new filtering period. 
 
The micro-controller can be placed in a calibration mode, where the system can be re-calibrated. 
In the calibration mode, the system slope is calculated from two points (zero and full scale) in the 
2-point  calibration  mode  or  the  slope  and  change  in  slope  is  calculated  from  three  points  (zero, 
half, and full scale) in the 3-point calibration mode. 
 

Содержание SR Scales SR585i-KG

Страница 1: ...CROBAN_201006 Page 1 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by...

Страница 2: ...ANTIMICROBIAL PROTECTION 8 STORAGE AND TRANSPORTATION 8 CLEANING AND DISINFECTING 9 SPECIFICATIONS 10 BUTTON FUNCTIONS 11 BASIC SYSTEM OPERATION 12 BATTERY REPLACEMENT 14 THEORY OF OPERATION 15 CALIBR...

Страница 3: ...n Scale DESCRIPTION QUANTITY Box 1 BASE ASSEMBLY 1 ea MAST ASSEMBLY with DISPLAY 1 ea DISPLAY UNIT WITH CABLE 1 ea HANDRAIL SET WITH HARDWARE 1 ea PACKAGE OF SIX 6 D CELL BATTERIES 1 ea MEDICAL GRADE...

Страница 4: ...splay cable through cable hole on the right STEP 3 Figure 1 Insert the four 4 3 8 16 screws 4 into the holes in Mast Assembly and Base Assembly as indicated Secure the Mast Assembly by tightening the...

Страница 5: ...le to the upright position Figure 2 Display Cable Connector and Battery Compartment STEP 10 Figure 3 Insert the Left 11 and Right 12 Side Hand Rails into the Rail Holder cups on the sides of the Base...

Страница 6: ...Bracket 9 to the rear of the Mast Assembly 1 using the two 2 mounting screws provided NOTE The three holes on the Printer Mast Bracket should be towards the top of the Mast Assembly figure 4 STEP 2 Ap...

Страница 7: ...flat head Mast Base Assembly mounting screws Item 4 Figure 1 to allow movement of the Mast for insertion of the Height Bar Assembly into the Base Assembly STEP 2 After Height Bar Assembly installation...

Страница 8: ...weight is displayed on a 16 character dot matrix LCD The weight data may be viewed in kilograms with a displayed resolution of 0 1 kg INTENDED USE The SR585i KG Bariatric Stand On Scale is specifical...

Страница 9: ...ct areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available...

Страница 10: ...ange AUTO ZERO One button operation AUTO POWER DOWN Approximately 30 seconds adjustable to 300 seconds RECALL Recalls last stored weight height and BMI AVERAGING Automatic digital filter POWER SUPPLY...

Страница 11: ...indicator flashes then remains solid when stable Auto stores stable weight in memory SEND PRINTER EHR Press to send stored values to EHR or to printer Output values include time date weight height and...

Страница 12: ...en PRINT INVERTED and PRINT NORMAL Press ENTER to save changes UP Press UP to adjust height up from the default to increase the scale s on time or to select a digit when setting time and date DOWN Pre...

Страница 13: ...ing a cane for support on the scale place the cane on the scale while zeroing the system This will ensure that the patient s NET weight will be displayed CALCULATING BMI Press to calculate BMI When th...

Страница 14: ...p the scale on its side and unscrew the four 4 screws in the Battery Compartment Cover located on the base of the unit between the wheels STEP 3 Remove and replace ALL six 6 D cell batteries Refer to...

Страница 15: ...ometers serving to normalize the sensitivity voltage out per unit of weight applied of each bridge The offset potentiometer produces a small current which nulls the output of the amplifier for an unlo...

Страница 16: ...ace the subset of calibrated weight on the scale Wait for scale to stabilize note scale reading STEP 4 Place all of the calibrated weight on scale Wait for scale to stabilize note scale reading Remove...

Страница 17: ...ed weights Use the UP arrow button to select the digit and the DOWN arrow button to change digit Press ENTER to save changes STEP 4 Press the MENU button until HALF 226 79 00 Kg is displayed Change HA...

Страница 18: ...system out of the Calibration Mode on the display board Figure 9 STEP 6 Place the display panel into the display housing Align the display panel with the four holes in the display housing Secure disp...

Страница 19: ...battery Wait five seconds then re install the battery and try the WEIGH button again The display lights appear to work but do not respond to button activation Check to ensure the membrane switch label...

Страница 20: ...or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles...

Страница 21: ...ng and Service Manual Part No MAN585i KG_MICROBAN_201006 Page 21 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstrumen...

Страница 22: ...o MAN585i KG_MICROBAN_201006 Page 22 of 22 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstr...

Отзывы: