6.2 Electromagnetic compatibility (EMC)
Other devices may interference even a slightly standard guidelines exceeding electromagnetic radiation
values. To check, if this bed is causing interference, stop use of this device by disconnecting it from mains and
check if it will make difference in other equipment. If malfunction in other device ends, this device may cause
noticed problems. This rare and unusual behavior can be reduced or eliminated by following methods:
•
Change position, distance or relocate compared to another equipment.
•
Ensure that used devices are suitable for existing enviroment.
Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the
other equipment should be observed to verify that they are operating normally
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the 4040X
examination table, including cables specified by the manufacturer. Otherwise, degradation of
the performance of this equipment could result.
Use of accessories, transducers and cables other than those specified or provided by the
manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
6.2.1 Electromagnetic emission
Medical device (Lojer 4040X) is intended for use in the electromagnetic environment specified below. The
customer or the user of the medical device should assure that it is used in such an environment.
Portable devices which are using radiofrequency can affect use of this equipment
Guidance and manufacturer’s declaration –
electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment-guidance
RF emissions
CISPR 11
Group 1, Class A Medical device uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment.
Harmonic
emissions IEC
61000-3-2
Class A
Device is directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Voltage fluctuations/ flicker
emissions IEC 61000-3-3
Complies