Copyright © Lojer Oy, 2021 | 05/2021
21
4 Cleaning and disinfecting
Before cleaning remove all accessories and unplug the power cord. Clean stains as soon as possible.
In order to keep the surfaces in good condition do the cleaning regularly. Do cleaning/disinfectant always
between patients. Do more thorough cleaning once a month. Follow the cleaning/disinfecting instructions given
by the respective facility.
Metal
–
and plastic surfaces
Clean the metal and plastic surfaces and the hand controls with a damp cloth
and weak alkaline cleaning fluid. Use small brush for corners and other difficult spots. Rinse with clean water
and dry carefully after cleaning. Do not use excessive fluids. Use disinfectant (alcohol or chlorine) and follow
the disinfectant manufacturer’s instructions fo
r use. Let dry by evaporation in room temperature.
Plastic surfaces (ABS, HDPE, PP)
are highly resistant to chemicals. Plastic is resistant to bleaching agents
(alkaline compounds), dilute organic or inorganic acids. Also solvents and cleaning agents may be used.
Plastic surfaces may get damaged if aromatic hydrocarbons (benzene and its derivates), ketones, ethers,
esters and chlorinated hydrogarbons are used. Plastic might also deteriorate if it is exposed to various
chemicals at the same time.
Stainless steel surfaces
are highly resistant to chemicals. Use for mild detergent solution for cleaning.
Ammonia and most of the solvents can be used to remove difficult stains. Avoid chlorine-based solutions.
Painted or chromed metal surfaces
can be cleaned with mild detergent. They are also highly resistant to
chemicals. Do not use harsh abrasive powders on these surfaces.
All surfaces must be dry before connecting to mains and using the device.
For safety reasons before cleaning unplug the power cord.
Do not use water spray (shower, high-pressure water guns) for cleaning.
Do not clean in high temperature and air humidity. Do not expose the device to excessive
moisture which can result in liquid pooling.
Do not use solvents or petrol for cleaning. Do not use acids for cleaning.
Disinfecting wears out the surfaces. After disinfecting clean the surfaces with clean, damp cloth.
Dilute the disinfectant according to t
he manufacturer’s instructions.
Please follow the instructions of the respective manufacturer when using common cleaners.
Do not use oil or grease based solutions.
Do not use chemical or dry cleaning on the material.
The material is not resistant to solvents, chlorides, washing/polishing agents or aerosol sprays.
Colorings (by jeans or other textiles) are excluded from any guarantee.
For hygienic reasons cover the upholstery with protective cloth or paper. Remove any stains as quickly as
possible with lukewarm water and a damp cloth. Microfiber cloth is recommended for this purpose. In case of
heavy soiling, use a mild cleaning agent and soft brush. Recommendable cleaning agent: Lojer Desiplint
(1:10), which is effective against bacteria without drying the upholstery material. Repeat the cleaning
procedure if necessary. (Composition of Lojer Desiplint: Chlorhexidine digluconate 0,1
–
0,2 %, water 99,8 %.)