II. Board Installation

Installing the RadioProcessor-G 

To install the board you must complete the following three steps:

1. Download and install the latest MRI SpinAPI software packages on your computer. MRI SpinAPI is 

available at: 

http://spincore.com/products/RadioProcessor-G/

MRI SpinAPI is a custom Application Programming Interface package developed by SpinCore

Technologies, Inc.  MRI SpinAPI is designed to be used only with the RadioProcessor-G.  

MRISpinAPI can be utilized using C/C++.

MRI SpinAPI can be supplemented with the 'MRI SpinAPI Examples' package.  A shortcut to 

the location of these files is created automatically in the Windows Start Menu at 

Start>Programs>SpinCore>MRI SpinAPI

”. 

2. Shut down the computer. Insert the RadioProcessor-G card into an available PCI slot and fasten the 

PC bracket securely with a screw.

3. Turn on the computer and follow the installation prompts.

4. In order to ensure the maximum accuracy of data acquisition for the RadioProcessor-G, the 

adc_offset.exe program needs to be run.  This program will produce a value that you will need to use 

to configure the ADC of the RadioProcessor-G, so please record this value for future reference.  For 

more details as to how to utilize the ADC Offset in a program, please refer to Section IV. Example 

Gradient Echo MRI Experiment  in this manual.

Testing the RadioProcessor-G

The simplest way to test whether the RadioProcessor-G has been installed properly and can be controlled 

as intended is to run a simple test program. 

Example_1.exe

 that accompanies the MRI SpinAPI Examples 

package (located at C:\SpinCore\MRI SpinAPI if the default installation directory was used) will produce a 

pulse with an on time and an off time of 500 ms on the two digital out BNCs with a duty cycle of 50%. This 

sequence will be repeated indefinitely until the program is exited by pressing any key after the board begins 

running. These outputs can be observed on an oscilloscope to confirm operation of the digital outputs.

Example_2.exe

 can be used to test the analog gradient outputs. Run the program and observe the three 

channels on an oscilloscope to check correct operation according to the program's description. This program 

will repeat until the program is exited by pressing any key. The board will be programmed one last time, 

triggered, and reset to zero the analog outputs.

After these two examples have been run, the 

2D_MRI.bat

 sample program can be run to test full 

functionality. This sample program can also be used to simply test the RF output. A description of the gradient 

echo pulse sequence and sample program can be found in the next section.

NOTE: There are additional programs in this folder. More information on these programs can be seen in the 

“Tips on obtaining data using the gradient echo example“ section.

2017-09-04

10

www.spincore.com

Содержание RadioProcessor-G

Страница 1: ...RadioProcessor G Owner s Manual SpinCore Technologies Inc http www spincore com...

Страница 2: ...ogies Inc reserves the right to make changes to the product s or information herein without notice RadioProcessor G PulseBlaster SpinCore and the SpinCore Technologies Inc logos are trademarks of Spin...

Страница 3: ...1 Step 2 Define RF Output Parameters 11 Step 3 Data Acquisition Parameters 12 Step 4 Specify Parameters of the Pulse Sequence 13 Step 5 Trigger the Pulse Program 14 Step 6 Retrieve Acquired Data 14 Us...

Страница 4: ...RadioProcessor G Hardware Trigger Reset Header JP601 20 Digital Output Header JP600 21 Related Products and Accessories 22 Contact Information 22 Document Information 22 2017 09 04 4 www spincore com...

Страница 5: ...essor manual for more details The system integrates SpinCore s high performance PulseBlaster timing engine for agile control of internal system components as well as TTL pulse pattern generation for c...

Страница 6: ...xcitation cores The gradient core controls three analog outputs that can be used to generate pulses with customizable durations and varying amplitudes of both positive and negative voltages A standard...

Страница 7: ...is shown below in Figure 2 2017 09 04 7 www spincore com Figure 2 Typical application of the RadioProcessor Model G By adding an RF power amplifier a gradient amplifier and a small signal pre amplifie...

Страница 8: ...it our website for more details Using the setup as described above with an 11 8 MHz permanent magnet a sample image of a 5 mm test tube filled with household cooking oil and two glass capillary tubes...

Страница 9: ...pling Frequency 75 1 MHz RF Analog Output D A Sampling Rate 300 MHz D A Sampling Precision 14 bits Output Voltage Range peak peak 1 2 2 V Phase resolution 0 09 deg Frequency resolution 0 28 Hz Gradien...

Страница 10: ...ent in this manual Testing the RadioProcessor G The simplest way to test whether the RadioProcessor G has been installed properly and can be controlled as intended is to run a simple test program Exam...

Страница 11: ...timing of the experiment i e pulse times delays etc 5 Trigger the pulse program The experiment will then proceed autonomously 6 Retrieve the captured data from the board at any time without interrupti...

Страница 12: ...nd in the RadioProcessor manual Once the RadioProcessor G has been instructed to acquire data there are two parameters that will influence the length for which data will be recorded these are the Spec...

Страница 13: ...e outputting data Multiple DACs can be selected int freq selects the frequency register that will be used for the carrier signal int tx_phase selects the phase register that will be used for the carri...

Страница 14: ...the SpinCore website includes a document that gives an in depth description of all the functions available in the API If you are unsure of how a particular function in an example program works please...

Страница 15: ...acquired per acquisition period nScans Number of times the entire sequence is performed nPhases Number of different phase levels to run the scan for spectrometerFrequency_MHz Frequency of the RF exci...

Страница 16: ...h all three gradients are applied Phase Gradient Time A period of time is waited to account for rise time of the gradient amplifiers Gradient Echo Delay The readout gradient coils then produce a gradi...

Страница 17: ...d gradient echo FID_Readout FID_Readout bat has delays and acquisition parameters set by default to aid in inspecting a single gradient echo This is best used to make sure that your acquisition parame...

Страница 18: ...outputs which can be controlled through the pulse program The analog output connector BNC3 is not equipped with an interpolating filter This allows for maximum flexibility in output frequency but it...

Страница 19: ...adout Gradient Hardware Trigger Reset Header JP601 The shrouded IDC header JP601 is the Hardware Trigger Reset connector This is an input connector for hardware triggering HW_Trigger and resetting HW_...

Страница 20: ...rigger is more precise since there are no software latencies involved HW_Reset pin 1 When this input is set to logical 0 for example by shorting it with pin 2 the pulse program is reset 2017 09 04 20...

Страница 21: ...nnector for header JP600 can be obtained on Digi Key Part Number 1 1658526 1 ND 2017 09 04 21 www spincore com Figure 9 Unshrouded IDC Output Header J600 Table 5 Unshrouded IDC Output Header J600 sign...

Страница 22: ...nent Magnets shtml 4 Consider complete MRI system in a small single bay enclosure Please inquire with SpinCore Technologies through our contact form which is available at http spincore com contact sht...

Отзывы: