Sper scientific AquaShock 850034K Скачать руководство пользователя страница 1

AquaShock® Water Purity Kit

850034K

Содержание AquaShock 850034K

Страница 1: ...AquaShock Water Purity Kit 850034K...

Страница 2: ...itted in any form or by any means electronic mechanical or other means that do not yet exist or may be developed including photocopying recording or any information storage and retrieval system withou...

Страница 3: ...ODUCTION 4 FEATURES 5 MATERIALS SUPPLIED 5 OPTIONAL ACCESSORIES 5 LCD DISPLAY 6 POWER SUPPLY 7 KEYPAD 8 SETUP MODE 9 CALIBRATION 14 MEASUREMENT PROCEDURES 17 SPECIFICATIONS 23 WARRANTY 24 TABLE OF CON...

Страница 4: ...Meter is fully shockproof floats and features a double injection molded case with an integral protective soft grip outer layer high end antishock LED backlight display and is powered by the latest env...

Страница 5: ...Fahrenheit selectable 99 data points memory for each measurement type Calibration point review Internal clock and calendar Maximum minimum and average Hold function Low battery indicator Backlight MA...

Страница 6: ...Scale Unit Current measurement scale unit 3 Measuring Results Current measurement value 4 ATC Automatic Temperature Compensation indicator 5 Temperature Scale Current temperature measurement unit 6 T...

Страница 7: ...charged More than 90 life remaining More than 85 90 life remaining More than 80 85 life remaining Less than 80 life remaining Probe Life Icon LCD DISPLAY POWER SUPPLY This meter is powered by one rech...

Страница 8: ...unction 2nd Function 1 Power Backlight Low Battery Power on off Backlight on off 2 Setup Setting parameters Confirm setting 3 MI Store data 4 MR Recall data 5 Max Min Ave Max Min Ave 6 Range Switch me...

Страница 9: ...on the LCD 4 Press or to select the C or F temperature unit 5 Press SETUP to save and return to Normal Mode 6 Press EXIT to return to Normal Mode without saving the selection Note When the temperatur...

Страница 10: ...st the month day hour and minute see Fig D E F 6 Press EXIT to return to Normal Mode without saving Note This procedure adjusts the meter s internal clock which is a 24 hour format only The internal c...

Страница 11: ...or NO see Fig C 5 When YES is displayed press SETUP to confirm and all parameters will restore to factory settings except the clock and the meter returns to Normal Mode 6 Press EXIT to return to Norm...

Страница 12: ...ode 6 Press EXIT to return to Normal Mode without saving change To calculate the temperature compensation coefficient 1 Set temperature coefficient to 0 2 Put conductivity probe into the standard cali...

Страница 13: ...e to confirm the TDS factor value and return to Normal Mode 6 Press EXIT once to return to Normal Mode without saving changes Fig A Fig B Cond value at 25 C TDS KCl TDS NaCl TDS 442 TDS Your Material...

Страница 14: ...atically select the correct range The measured value is displayed above the line the set calibration value is shown flashing below the line see Fig A 2 Press once to increase the Calibration value Hol...

Страница 15: ...alue is shown flashing below the line see Fig A 2 Press once to increase the calibration value Hold down to continually increase the value see Fig B Press once to reduce the Calibration value Hold dow...

Страница 16: ...e is a cyclic display the Conductivity scale will display as in Fig B and Fig C The TDS scale will display as in Fig D and Fig E If there are no calibration points set the display will show see Fig F...

Страница 17: ...cm scales and will calculate the TDS value in ppm or ppt according to the TDS conversion coefficient that was entered in the TDS Factor set up procedure During measurement and calibration this meter w...

Страница 18: ...seconds to turn the meter off Indicator Light Descriptions Meter Status Indicator Status Power off Light off Power on Green light on for 5 seconds then light off Low battery Red light flashing every...

Страница 19: ...ve the current measured value and view the data point number see Fig A 2 Up to 99 Memories can be saved for both Conductivity and TDS If the saved memory exceeds 99 the new data will be written over s...

Страница 20: ...ng see Fig A and Fig B The date and time are on cyclical display If there is no saved data is displayed see Fig C 3 Press or to view the saved data points Press to view the data points increasing by 1...

Страница 21: ...see Fig A The meter will return to Normal mode Hold 1 Press HOLD to hold the value see Fig A 2 Press HOLD to release the hold value Note When in Hold Mode all other functions will be disabled except...

Страница 22: ...layed see Fig C 3 Press MAX MIN AVE the minimum recorded value will appear on the LCD see Fig D and Fig E The date and time are on cyclical display If no measurements have been taken is displayed see...

Страница 23: ...0 01 S 0 1 S 1 S 0 01 mS 0 1 mS 0 01 ppm 0 1 ppm 1ppm 0 01 ppt 0 1 ppt Accuracy 0 5 C 0 9 F 1 full scale 1 full scale Input Temperature Probe Connector BNC Connector Calibrate 5 Points 1 point in each...

Страница 24: ...not cover probes batteries battery leakage or damage resulting from accident tampering misuse or abuse of the product Opening the meter to expose its electronics will break the waterproof seal and voi...