The front pull-out handles can preferably be pulled out or closed.
•
The siderails of the stretcher must be raised.
•
The trolley parking brakes (normal and intermediate position) must be engaged or other operators must ensure that the trolley is
stationary.
•
Check that the trolley bed is free of any foreign objects (hands, sheets, accessories).
•
Lift the stretcher and place it on the trolley without altering the previously defined alignment.
•
Check again that no operators have inadvertently placed their hands anywhere on the frame between the stretcher and the trolley, as
there is a
hazard of crushing fingers
.
•
Once you have checked that the stretcher is correctly centred on the trolley and that the wheel supports are resting on the trolley slide
profile, you can push the stretcher forward until the locking mechanism engages.
•
Make sure that no fabric or accessories have interfered with insertion of the locking mechanism between the stretcher and the trolley.
•
Close the pull-out handles before moving the trolley and stretcher assembly.
11.19
Restraint straps
To apply the two-piece restraint straps, identify a portion of the frame at chest height and one at the patient's leg level. The restraint straps
should be applied in a position that allows for adequate immobilization of the patient.
Secure to the frame by means of a slipknot
–
Image P51
The restraint straps supplied with the stretcher must be fastened immediately after positioning the patient.
To do this, insert the male buckle into the female buckle
–
Image P52
When using the chest strap, the male buckle must be passed through the appropriate slots in the metal ends of the chest straps
–
Image P53
.
Tension the straps after attaching them
–
Images P54 and P55
When the patient is to be moved from the stretcher at the end of use, release the straps by pressing the release button marked "PRESS"
–
Images P56 and P57.
Then open the stretcher siderail on the side towards which you want to lower the patient from the stretcher.
DNA STRAP:
DNA straps are suitable for use with stretchers equipped with a patient bed with special housing.
Follow the instructions below to apply them, if they have not already been installed at the factory
–
Images P58, P59, P60, P61
:
1
–
Remove the strap from the seat present on the re-winding system.
2
–
Insert the DNA system, keeping the belt taut, into the special cavity on the body of the bed. The male and female straps must be on the
left and right of the platform, respectively.
3
–
Put the belts back into the hole on the DNA re-winding system and complete the installation.
4
–
Proceed in the same way for thorax systems.
After positioning the patient, fasten the straps and check that they are correctly applied.
DNA straps are not equipped with a pre-tensioner. It is therefore necessary to manually adjust them after fastening them, tensioning the
straps and bringing the re-winding system to the end of the stroke.
Incorrect adjustment makes the device unsafe and ineffective.
12.
CLEANING AND MAINTENANCE
Spencer Italia S.r.l. declines all responsibility for any direct or indirect damage which is the consequence of improper use of the product
and spare parts and/or in any case of any repair carried out by a person other than the Manufacturer, who uses internal and external
technicians authorised to do so; moreover, doing so will invalidate the warranty.
•
The operator must wear suitable personal protective equipment, such as gloves, goggles, etc. during all checking, maintenance and
cleaning operations.
•
Establish a maintenance schedule, periodic inspections and extend the average life span, if foreseen by the Manufacturer in the User
Manual, identifying a reference person who meets the basic requirements set forth in the User Manual.
•
The frequency of inspections is determined by factors such as legal requirements, type of use, frequency of use, and environmental
conditions during use and storage.
•
Repairs of products manufactured by Spencer Italia S.r.l. must be carried out by the Manufacturer, who shall make use of specialised
internal or external technicians who, using original spare parts, shall provide quality repair service in strict compliance with the
technical specifications indicated by the Manufacturer. Spencer Italia S.r.l. declines any responsibility for any direct or indirect damage
which is a consequence of improper use of spare parts and/or any repair work carried out by unauthorised parties.
•
Use only original or Spencer Italia S.r.l. approved components/spare parts and/or accessories in order to carry out all operations without
causing alterations or modifications to the product.
•
All maintenance and overhaul activities must be recorded and documented with the relevant technical operation reports. This
documentation must be kept for at least 10 years after the end of the product's life and must be made available to the competent
authorities and/or the Manufacturer when requested.
•
Cleaning, provided for reusable products, must be carried out in accordance with the Manufacturer's instructions in the User Manual in
order to avoid the risk of cross-infection due to the presence of body fluids and/or residues.
•
The product and all its components must be washed and left to dry completely before storage.
•
If the product requires lubrication, this must be done after cleaning and complete drying.
•
The dedicated stretcher fixing system is equipped with a system that signals the need for stretcher maintenance and of the fixing itself
linked to the cycles of use. Strictly observe the required maintenance intervals.
•
Document maintenance using the form in Annex B of this manual or other appropriate forms.
Содержание CROSS
Страница 2: ...2 P1 P2 P3 P4 P5 ...
Страница 3: ...3 P6 P7 P8 P9 P10 P11 P12 P13 P14 ...
Страница 4: ...4 P15 P16 P17 P18 P19 P20 P21 P22 ...
Страница 5: ...5 P23 P24 P25 P26 P27 P28 P29 P30 P31 P32 SHARP SHARP SL P33 P34 ...
Страница 6: ...6 P35 P36 P37 P38 P39 P40 P41 P42 P43 P44 P45 P46 ...
Страница 7: ...7 P47 P48 P49 P50 P51 P52 P53 P54 P55 P56 P57 ...
Страница 8: ...8 P58 P59 P60 P61 ...