The front pull-out handles can preferably be pulled out or closed.

 

 

The siderails of the stretcher must be raised. 

 

The  trolley  parking  brakes  (normal  and  intermediate  position)  must  be  engaged  or  other  operators  must  ensure  that  the  trolley  is 
stationary.

 

 

Check that the trolley bed is free of any foreign objects (hands, sheets, accessories).

 

 

Lift the stretcher and place it on the trolley without altering the previously defined alignment. 

 

Check again that no operators have inadvertently placed their hands anywhere on the frame between the stretcher and the trolley, as 
there is a 

hazard of crushing fingers

 

Once you have checked that the stretcher is correctly centred on the trolley and that the wheel supports are resting on the trolley slide 
profile, you can push the stretcher forward until the locking mechanism engages.  

 

Make sure that no fabric or accessories have interfered with insertion of the locking mechanism between the stretcher and the trolley. 

 

Close the pull-out handles before moving the trolley and stretcher assembly. 

11.19

 

Restraint straps 

To apply the two-piece restraint straps, identify a portion of the frame at chest height and one at the patient's leg level. The restraint straps 
should be applied in a position that allows for adequate immobilization of the patient. 
Secure to the frame by means of a slipknot 

 Image P51

 

The restraint straps supplied with the stretcher must be fastened immediately after positioning the patient. 
To do this, insert the male buckle into the female buckle  

 Image P52 

When using the chest strap, the male buckle must be passed through the appropriate slots in the metal ends of the chest straps 

 Image P53

Tension the straps after attaching them  

 

Images P54 and P55 

When the patient is to be moved from the stretcher at the end of use, release the straps by pressing the release button marked "PRESS" 

 

Images P56 and P57. 

Then open the stretcher siderail on the side towards which you want to lower the patient from the stretcher. 

DNA STRAP: 

DNA straps are suitable for use with stretchers equipped with a patient bed with special housing. 
 
Follow the instructions below to apply them, if they have not already been installed at the factory 

 

Images P58, P59, P60, P61

 Remove the strap from the seat present on the re-winding system. 

 Insert the DNA system, keeping the belt taut, into the special cavity on the body of the bed. The male and female straps must be on the 

left and right of the platform, respectively. 

 Put the belts back into the hole on the DNA re-winding system and complete the installation. 

 Proceed in the same way for thorax systems. 

After positioning the patient, fasten the straps and check that they are correctly applied. 
DNA straps  are not equipped with a pre-tensioner. It  is therefore necessary to manually adjust  them after fastening them, tensioning the 
straps and bringing the re-winding system to the end of the stroke. 
Incorrect adjustment makes the device unsafe and ineffective. 

12.

 

CLEANING AND MAINTENANCE 

Spencer Italia S.r.l. declines all responsibility for any direct or indirect damage which is the consequence of improper use of the product 
and spare parts and/or in any case  of any repair  carried  out by a  person other than the Manufacturer, who  uses internal and external 
technicians authorised to do so; moreover, doing so will invalidate the warranty. 

 

The  operator  must  wear  suitable  personal  protective  equipment,  such  as  gloves,  goggles,  etc.  during  all  checking,  maintenance  and 
cleaning operations. 

 

Establish  a  maintenance  schedule,  periodic  inspections  and  extend  the  average  life  span,  if  foreseen  by  the  Manufacturer  in  the  User 
Manual, identifying a reference person who meets the basic requirements set forth in the User Manual. 

 

The frequency of inspections is determined  by factors such as legal requirements, type of  use,  frequency of  use, and environmental 
conditions during use and storage.  

 

Repairs  of  products  manufactured  by  Spencer  Italia  S.r.l.  must  be  carried  out  by  the  Manufacturer,  who  shall  make  use  of  specialised 
internal  or  external  technicians  who, using  original  spare  parts,  shall  provide      quality  repair  service    in  strict  compliance  with  the 
technical specifications  indicated by the Manufacturer. Spencer Italia S.r.l. declines any responsibility for any direct or indirect damage  
which is a consequence of improper use of spare parts and/or any repair work carried out by unauthorised parties. 

 

Use only original or Spencer Italia S.r.l. approved components/spare parts and/or accessories in order to carry out all operations without 
causing alterations or modifications to the product. 

 

All  maintenance  and  overhaul  activities  must  be  recorded  and  documented  with  the  relevant  technical  operation  reports.  This 
documentation  must  be  kept  for  at  least  10  years  after  the  end  of  the  product's  life  and  must  be  made  available  to  the  competent 
authorities and/or the Manufacturer when requested. 

 

Cleaning, provided for reusable products, must be carried out in accordance with the Manufacturer's instructions in the User Manual in 
order to avoid the risk of cross-infection due to the presence of body fluids and/or residues. 

 

The product and all its components must be washed and left to dry completely before storage. 

 

If the product requires lubrication, this must be done after cleaning and complete drying. 

 

The dedicated stretcher fixing system is equipped with a system that signals the need for stretcher maintenance and of the fixing itself 
linked to the cycles of use. Strictly observe the required maintenance intervals. 

 

Document maintenance using the form in Annex B of this manual or other appropriate forms. 

Содержание CROSS

Страница 1: ...PENCER ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy SPENCER ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy www spencer it support spencer it e mail info spencer it service service spencer it Ph Tel 39 0521 541111 Quality System EN ISO 9001 EN ISO 13485 ...

Страница 2: ...2 P1 P2 P3 P4 P5 ...

Страница 3: ...3 P6 P7 P8 P9 P10 P11 P12 P13 P14 ...

Страница 4: ...4 P15 P16 P17 P18 P19 P20 P21 P22 ...

Страница 5: ...5 P23 P24 P25 P26 P27 P28 P29 P30 P31 P32 SHARP SHARP SL P33 P34 ...

Страница 6: ...6 P35 P36 P37 P38 P39 P40 P41 P42 P43 P44 P45 P46 ...

Страница 7: ...7 P47 P48 P49 P50 P51 P52 P53 P54 P55 P56 P57 ...

Страница 8: ...8 P58 P59 P60 P61 ...

Страница 9: ...PERIODIC OVERHAUL 16 12 4 SPECIAL MAINTENANCE 17 12 5 LIFE SPAN 17 13 TROUBLESHOOTING TABLE 17 14 ACCESSORIES 18 14 1 ACCESSORIES 18 15 SPARE PARTS 18 16 DISPOSAL 19 Notice The information contained in this document is subject to change without notice and is to be intended as a commitment by Spencer Italia S r l subject to change The images are included as examples and may vary slightly from the a...

Страница 10: ...best instructions are the continuous practice of use under the supervision of competent and trained personnel Regardless of your level of experience with similar devices in the past you should carefully read and understand the contents of this manual before installing operating or servicing this product In case of any questions please contact Spencer Italia S r l for the necessary clarifications T...

Страница 11: ...be traced the device must be reconditioned provided only under the responsibility of the manufacturer Regulation 2017 745 EU requires manufacturers and distributors of medical devices to keep track of their location If the device is in a location other than the address to which it was shipped or sold or if it was donated lost stolen exported or destroyed permanently removed from use or if the devi...

Страница 12: ...nts Avoid contact with sharp objects Device installation must be carried out by qualified personnel trained and enabled by Spencer Italia S r l The times and methods of such shall be agreed upon between the customer and our Sales Offices Operating temperature from 10 C to 50 C Storage The product must not be exposed or come into contact with thermal sources of combustion or flammable agents but mu...

Страница 13: ...n any way replace his or her supervision Pay the utmost attention to any obstacles water ice debris etc present on the route as they may cause the operator to lose balance and compromise proper functioning of the device If you cannot clear the way choose an alternative route Condensation water ice and accumulations of dust can affect correct operation of the device making it unpredictable and caus...

Страница 14: ...y result in injury to the patient rescuer or third parties Inadequate disinfection procedures may lead to risk of cross infection Partial opening of the legs may result in the device falling to the ground Make sure that the legs are properly locked before any handling and that the pistons are fully extended and stable If the stretcher is not locked on the fixing system or if it is not correctly po...

Страница 15: ... 19 Front legs Steel 12 Rear lock system Steel Nylon Aluminium 20 Variable height control lever Nylon 13 Rear wheel with bracket PU 21 Loading trolley wheel PU 14 Brake Chrome plated FE 22 Stops for Cross Up 8409 stretcher Steel Nylon 15 Rear leg lifting piston Steel 23 Front wheel unlocking lever TWIST version only Polyethylene 16 Rear legs Steel 24 Stretcher release lever Nylon 17 Front leg lift...

Страница 16: ...nt bed height with trolley at intermediate height mm 760 830 Cross Chair patient bed height with trolley lowered mm 395 Leg and loading trolley wheels mm Ø 200 Load capacity kg 170 250 Crossover Chair Weight kg 32 Available in 10g version Note 1 The patient bed height specifications are measured without mattress Note 2 The double dimensional values in some fields refer to the rear and front of the...

Страница 17: ...ng on whether you wish to increase or reduce the loading height The new position must allow for insertion of both previously removed screws P11 Insert the screws and partially tighten them Carry out the same operation on the plate located on the opposite side of the stretcher P12 Move the loading trolley to its new position Place the loading trolley on the vehicle floor and check that the height i...

Страница 18: ...king handle green When engaged it unlocks the movement of the rear leg allowing it to be closed E Front leg unlocking handle red When engaged it unlocks the movement of the front leg allowing it to be closed F Rear wheel brake When engaged it brakes the rear wheels G Sidebar unlocking levers When engaged at the same time they unlock the sidebar and open it It is not necessary to operate any contro...

Страница 19: ...om the ground during loading unloading or when the loading trolley is still on the vehicle to allow the front leg to open safely Failure to take the above measure will preclude safe use of the device resulting in risk of damage to the patient operators and the device itself 11 2 Parking brakes Image P20 To engage the parking brakes simply press on the strips on the rear wheel brackets with your fo...

Страница 20: ...g the stretcher from the trolley Cross Up and Cross Chair Images P27 and P28 Brake the trolley with the parking brakes Engage the corresponding control lever and pull out the stretcher sliding it about ten centimetres Complete the operation by lifting the stretcher from the trolley This operation requires at least two operators one on the foot side of the stretcher to operate the controls one on t...

Страница 21: ...n To reach this position lift the end of the board by grasping the two rings at its end Ensure that the set screw is properly seated in its socket before releasing the rings To change from the Trendelenburg position to the standard position lift the patient bed slightly by grasping the end of the frame check that the set screw is out of its socket and then support the frame as it goes back down to...

Страница 22: ...able part of the frame as it moves towards the ground Caution Hazard of crushing fingers Make sure that your hands or those of other operators or other objects are not in the area where the mobile frame joins the fixed frame highlighted in the figure with the warning symbol Return the rear telescopic handles to Position 1 The patient can only be lifted after checking that these are fully inserted ...

Страница 23: ... not equipped with a pre tensioner It is therefore necessary to manually adjust them after fastening them tensioning the straps and bringing the re winding system to the end of the stroke Incorrect adjustment makes the device unsafe and ineffective 12 CLEANING AND MAINTENANCE Spencer Italia S r l declines all responsibility for any direct or indirect damage which is the consequence of improper use...

Страница 24: ...e device The operator must wear suitable personal protective equipment such as gloves goggles etc during all checking maintenance and cleaning operations Checks to be carried out before and after each use and at deadline indicated above are as follows General functionality of the device Cleanliness of the device remember that the failure of cleaning may cause the risk of cross infections Correct t...

Страница 25: ...e the device from service and contact the service centre The stretcher is not correctly attached to the fixing system Wear or damage of the components that constitute the locking mechanisms Immediately remove the device from service and contact the service centre The fifth wheel has not been properly inserted in the fixing system Position the stretcher properly and check that the fifth wheel enter...

Страница 26: ...702C Hook R MAX B certified 10G IF01047C TRACK 4 30 TELESCOPIC IV HOLDER EN90003C END T 10G INSTRUMENTS TABLE CB09028C TANKER FIXING SUPPORT FOR SAFEBAR CB09025C TANKER YELLOW OXY TANK HOLDER FOR STRETCHER SIDERAIL FIXING ST00491A STX 90 TELESCOPIC HEADREST FOR STRETCHERS CR90010B Adapter for loading trolley wheels CR90011B S MAX HUX LOADING TROLLEY EXTENDER CR90012C LOADING TROLLEY ADAPTER FOR S ...

Страница 27: ...ing system ST00498B DNA Strap with re winding system 16 DISPOSAL When devices and their accessories are no longer suitable for use they can be disposed of as normal municipal solid waste if they have not been contaminated by special agents Otherwise follow the regulations in force regarding disposal ...

Отзывы: