10 

 

See image P7 -

 Stretchers intended for use with the Sensor Lock fixing system, identified as "SL", are equipped with a communication system 

with the fixing system installed on the fifth wheel of the stretcher. Before putting into service, check that the unlocking system is working 
properly. In the event of a failure following the adjustments described above, it may be necessary to adjust the positioning of this system by 
moving the magnet in order to restore proper function. 
The positioning of this system must be evaluated on site and following the stretcher adjustments described above. 
 
 

MEASURING  OF  THE  LOADING  FLOOR  MUST  BE  CARRIED  OUT  AFTER  HAVING  SET  THE  AMBULANCE  ON  AN  EVEN  SURFACE  WITH  NO 
BUMPS  AND/OR  UNEVENNESS  AND  AFTER  ACCOMMODATING  2  (TWO)  OPERATORS  AT  THE  REAR  OF  THE  SANITARY  UNIT,  WITH  A 
SIMULATED LOAD OF UP TO 250 KG. 
If the vehicle is equipped with pneumatic or hydraulic suspensions, the loading  height must be adjusted taking into account the worst-
case conditions and/or the working conditions foreseen by the outfitter. 
Any problems regarding the use and/or safety risks associated with this system are not the responsibility of the manufacturer.   

 
 
 
 

10.

 

OPERATING CHARACTERISTICS 

CROSS AND CROSSOVER 

 

See image 

P8 

 P9 

 P10

 

 

Element 

Description 

Function 

Patient bed 

In Trendelenburg or Trendelenburg Fowler version, it is the area on which 
the patient rests and is equipped with an adjustable backrest 

Leg unlocking levers 

Red  for  the  front  legs  and  green  for  the  rear  legs;  unlocks  movement  of 
the respective legs 

Fifth wheel 

Element for coupling with the rear part of the fixing system 

Brakes 

Placed on the rear wheel brackets to lock wheel rolling and pivoting 

Variable height unlocking lever 

Only  available  on  XL  variants,  allowing  the  stretcher  to  be  positioned  at 
intermediate height 

Sidebar unlocking lever 

Unlock sidebar opening when right and left are pressed at the same time 

Loading trolley unlocking sleeve 

Only available on ACTIV versions; unlocks movement of the loading trolley 
when activated, allowing the stretcher to be shortened 

Back lift piston knob 

Pulling it releases the stop to allow the backrest to be returned to a lower 
position than it is currently in 

Telescopic handle unlocking buttons 

Only available on the TH version; unlocking telescopic handle sliding 

11.

 

PROPER USE 

Primary medical evaluations must be carried out before the patient is moved, lifted or transported. Once the diagnosis has been made, it is 
preferable to advise the patient to actively contribute to the transition from bed to stretcher/chair, while at the same time informing them of 
the risks they may be running. Before loading the patient, bring the device as close as possible to the patient. 

11.1

 

Emergency vehicle requirements 

The stretcher is designed to enter and exit the medical compartment of an ambulance. The vehicle requirements should be:  

 

Level stretcher floor 

 

Stretcher floor that is wide and long enough to accommodate the stretcher without becoming an obstacle  

The wheels of the front legs must be at a safe distance of at least 5 cm from the ground during loading/unloading or when the loading trolley 
is still on the vehicle, to allow the front leg to open safely 

Failure to take the above measure will preclude safe use of the device, resulting in risk of damage to the patient, operators and the device 
itself. 

11.2

 

Parking brakes - Image P12 

To engage the parking brakes, simply press on the strips on the rear wheel brackets with your foot. To disengage the parking  brakes, simply 
press down on the opposite side of the parking brake pedal and they will return to their original position with a slight click. 
Never leave the patient unattended, even if the parking brakes have been engaged. 

11.3

 

Sidebars - Image P13 

The stretcher is equipped with sidebars, which are necessary for restraining the patient on the stretcher.

  

 
 
 

 

Содержание Carrera Tec

Страница 1: ... ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy SPENCER ITALIA SRL Via Provinciale n 12 43038 Sala Baganza PR Italy www spencer it support spencer it e mail info spencer it service service spencer it Ph Tel 39 0521 541111 Quality System EN ISO 9001 EN ISO 13485 ...

Страница 2: ...2 P1 P2 P3 P4 ...

Страница 3: ...3 P3 P4 P5 P6 P7 P8 P9 P10 ...

Страница 4: ...4 P12 P13 P14 P15 P16 SHARP SHARP SL P17 P18 P19 P20 P21 P22 P23 ...

Страница 5: ...5 P24 P25 P26 P27 ...

Страница 6: ... P14 and P15 11 11 5 Trendelenburg Board Image P17 11 11 6 Trendelenburg Fowler Board Image P18 11 11 7 Backrest adjustment Image P19 12 11 8 Changing the stretcher length ACTIV only Image P20 P21 12 11 9 Lowering the stretcher Image P22 P23 12 11 10 Telescopic handle operation Image P26 13 11 11 Lifting the stretcher with the patient 13 11 12 Spine board housing Carrera Pro only Image 27 13 12 CL...

Страница 7: ...Rev 0 04 05 2021 Code CCI5297 Copyright Spencer Italia S r l All rights reserved No part of the document may be photocopied reproduced or translated into another language without prior written consent from Spencer Italia S r l ...

Страница 8: ...instructions are the continuous practice of use under the supervision of competent and trained personnel Regardless of your level of experience with similar devices in the past you should carefully read and understand the contents of this manual before installing operating or servicing this product In case of any questions please contact Spencer Italia S r l for the necessary clarifications The pr...

Страница 9: ...y will be void as the device can no longer be traced If the assigned Lot SN cannot be traced the device must be reconditioned provided only under the responsibility of the manufacturer Regulation 2017 745 EU requires manufacturers and distributors of medical devices to keep track of their location If the device is in a location other than the address to which it was shipped or sold or if it was do...

Страница 10: ...ht which weighs on each operator must comply with local health and safety requirements Avoid contact with sharp objects Device installation must be carried out by qualified personnel trained and enabled by Spencer Italia S r l The times and methods of such shall be agreed upon between the customer and our Sales Offices Operating temperature from 10 C to 50 C Storage The product must not be exposed...

Страница 11: ...e utmost attention to any obstacles water ice debris etc present on the route as they may cause the operator to lose balance and compromise proper functioning of the device If you cannot clear the way choose an alternative route Condensation water ice and accumulations of dust can affect correct operation of the device making it unpredictable and causing a sudden change in the weight the operators...

Страница 12: ...sual circumstances the presence of multiple operators is recommended not just 2 as expected in standard conditions The ability of each operator must be assessed before the definition of rescuer roles in use of the stretcher 7 RESIDUAL RISK The residual risks listed below have been identified solely with reference to the intended use of the device Use by untrained personnel may result in injury to ...

Страница 13: ... fixing without intervening on the fixing itself Nylon PE Steel 8 Front leg When closed it allows the stretcher to be loaded into the ambulance two wheels are attached to the stretcher which are necessary for its movement Steel 23 Sharp system drive handle if present Control part of the Sharp System Nylon Steel 9 Front wheel needed to move the stretcher PU 24 TEC TH Rear leg unlocking lever Nylon ...

Страница 14: ...operation and damage to the patient or rescuers and will void the warranty and release the Manufacturer from all liability Carrera stretchers are available in various heights to be assessed before purchase in relation to the loading configuration on the vehicle NOTE Please contact the manufacturer for installations requiring the use of accessories The type of stretcher should be chosen taking into...

Страница 15: ...ng lever Unlock sidebar opening when right and left are pressed at the same time G Loading trolley unlocking sleeve Only available on ACTIV versions unlocks movement of the loading trolley when activated allowing the stretcher to be shortened H Back lift piston knob Pulling it releases the stop to allow the backrest to be returned to a lower position than it is currently in I Telescopic handle unl...

Страница 16: ...AL STRETCHER MOVEMENTS NEVER ENGAGE THE LEG LOCKING LEVERS AS DOING SO COULD CAUSE THE STRETCHER TO FALL TO THE GROUND THESE LEVERS MUST ONLY BE ENGAGED TO LOAD THE STRETCHER ON THE AMBULANCE OR TO REACH THE INTERMEDIATE HEIGHT AS DESCRIBED ABOVE To unload the stretcher from the medical vehicle If the stretcher is used with the Sensor Lock fixing system release the stretcher by engaging the rear l...

Страница 17: ...d patients onto the stretcher at intermediate height Engage the rear leg unlocking lever hold it in place and lift the stretcher slightly to unlock the rear leg moving mechanism Lower the stretcher about 10 cm release the unlocking lever and support the stretcher until the intermediate height position is reached Make sure that the stretcher has reached a stable position Engage the brakes on the re...

Страница 18: ...injury to the patient or the operator by applying excessive force on the device 12 CLEANING AND MAINTENANCE Spencer Italia S r l declines all responsibility for any direct or indirect damage which is the consequence of improper use of the product and spare parts and or in any case of any repair carried out by a person other than the Manufacturer who uses internal and external technicians authorise...

Страница 19: ...ce maintenance must be carried out by operators with specific qualifications training and instruction in the use and maintenance of the device The operator must wear suitable personal protective equipment such as gloves goggles etc during all checking maintenance and cleaning operations Checks to be carried out before and after each use and at deadline indicated above are as follows General functi...

Страница 20: ...g levers don t work properly or are difficult to operate The moving mechanisms are damaged Immediately remove the device from service and contact the service centre Means of connection between components have been lost Immediately remove the device from service and contact the service centre The stretcher is not correctly attached to the fixing system Wear or damage of the components that constitu...

Страница 21: ...faulty Check the fixing system power supply If the problem persists immediately remove the device from service and contact the service centre The stretcher is used with a fastener that does not provide automatic release If necessary request a different fixing system ...

Страница 22: ...ow oxygen tank for stretcher siderail fixing CB09026C Siderail fixing support for tanker oxygen tank ST00497B DNA Strap chest strap with re winding system ST00498B DNA Strap with re winding system ST00499B STX 499 UNIVERS ADJUST 4 PC CHEST STRAP ST00592A STX 592 YELLOW METAL 2 PC METAL HOOK STRAP 15 SPARE PARTS ST70002A STX 702 Two piece metal black reflex strap black ST70000A QMX 777 Anatomical s...

Отзывы: